Introduction 

 

CAUTION 

Federal Law in the U.S.A. and Canada restricts this device to sale by or on the order of a licensed practitioner. 

 

Intended Use 

The cabinet mount installation kit and accessories are 
intended for use in mounting the Digital MDM® mixer 
and associated accessories. The Digital MDM mixer is 
intended for use by trained medical personnel only. 
 

User and Installer Responsibility 

It is the responsibility of both the user and the installer 
to ensure that the correct gas, at the correct pressure, 
is supplied to the fittings at the rear of the unit. See 
Specifications in the Digital MDM Operations Manual. 
 

System Notes 

Matrx by Parker recommends the use of a permanently 
mounted oxygen analyzer in the patient circuit to 
monitor the oxygen levels being administered to the 
patient. Additional monitoring should be employed 
particularly if the delivered oxygen and nitrous oxide is 
being used as background to intravenous or inhalant 
deep sedation and anesthesia. TheAmerican Society of 
Anesthesiology (ASA) encourages the use of an ECG, 
a Blood Pressure Monitor, and a Pulse Oximeter in 
these cases. 
 

NFPA Requirements and Warnings 

 
General System Requirements 
The Digital MDM mixer complies with the National Fire 
Protection Association (NFPA) 99C code requirements. 
 
The NFPA is a body which has formulated a set of 
regulations widely adopted by local building authorities 
as the basis for their own local building code 
requirements. 

Do not change fittings or tubing sizes 

from those specified in this manual

. The fittings and 

tubing incorporate unique tubing sizes for each gas, an 
important safety feature to prevent accidental crossing 
of lines during installation or maintenance. 
 
The following basic requirements must be observed: 
 
1.  Remote tank rooms must be locked. 
 
2.  Ventilation of at least 72 square inches is required 

in the tank room or the cabinet. 

  
 
 
 
 

 
 
 

WARNING 

Medical gas fittings are manufactured according to 
the Diameter Indexed Safety System (DISS) to 
prevent the delivery of the wrong gas to the patient. 
The correct supply line sizes are as follows: 

oxygen  

1/2” OD copper 

nitrous oxide  

3/8” OD copper  

evacuation 

5/8” OD copper 

The installer must ensure that each gas is 
connected to the proper stop valve to preserve the 
integrity of the Diameter Indexed Safety System 
(DISS). 

 
3.  Type K or L seamless, copper tubing or standard 

weight brass piping is required. 

 
4.  Copper-to-copper joints made using phosphorous 

brazing filler are allowed without flux. 

 
5.  During brazing, piping must be continuously purged 

with dry nitrogen. 

 
6.  The piping system must be tested for leaks at 150 

psi followed by a 24-hour pressure test at 75 psi. 

 

NOTE 

Do not connect the Digital MDM during these 
pipeline tests. 

 

Manifold Requirements 

The Digital MDM mixer is intended for Type II facilities 
consisting of less than 2000 cu.ft. of gas and less than 
ten use points for the gas. 
 
1.  All systems shall be a minimum 2+2 configuration. 
 
2.  For a non-remote tank room, manual or automatic 

switchover is permitted. For a remote tank room, 
automatic switchover is required. 

 

NOTE 

A tank room is considered remote when it 
cannot be accessed from within the office. 

 
3.  A remote tank room requires a remote shutoff 

valve. 

 
4.  Each tank requires independent pressure 

regulators. 

 

 

Содержание Matrx Digital MDM

Страница 1: ...Ref 10550800 Rev F Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA Tel 215 723 4000...

Страница 2: ...g addresses Porter Regulatory Affairs EU Authorized Representative Parker Hannifin Corporation Parker Hannifin Ltd Porter Instrument Instrumentation Products Division 245 Township Line Road Riverside...

Страница 3: ...ility A licensed electrical person should be consulted to confirm the Protective Earth ground connection Do not use an electrical cord with nicks cuts or other damage Do not use the unit with an exten...

Страница 4: ...enging Control Valve Installation 6 1 6 J Hook Installation 8 1 7 Hose Connection 8 1 8 Rubber Goods Installation 9 1 9 Installation of Power Cord and Protective Earth ground Connection 10 Section 2 S...

Страница 5: ...ns 5 Figure 4 Screw and Tee Nut Adjustment 5 Figure 5 Digital MDM Mixer Head Installation 6 Figure 6 Flush Mounting the Bag Tee and Scavenging Control Valve 6 Figure 7 Surface Mounting the Bag Tee and...

Страница 6: ...tact Information Unit Serial Number Date Purchased Dealer Name Dealer Phone Number Notes Contact Information Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA...

Страница 7: ...he fittings and tubing incorporate unique tubing sizes for each gas an important safety feature to prevent accidental crossing of lines during installation or maintenance The following basic requireme...

Страница 8: ...information on the NFPA 99C code requirements consult your local Matrx by Parker representative or your local dental dealer U S Gas Color Coding All products in this manual are manufactured according...

Страница 9: ...A guideline No 99C see the Introduction for details All equipment must be installed in accordance with these guidelines as well as all local codes and ordinances Matrx by Parker makes no warranty for...

Страница 10: ...tion 1 Install the necessary gas plumbing from the central gas supply 2 If separate gas outlet stations are to be installed install the necessary gas plumbing See the Outlet Stations Installation and...

Страница 11: ...4 mounting holes See Figure 4 6 While supporting the front of the cabinet tap in the tee nuts with a hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certa...

Страница 12: ...installed The attachment to the cabinet surface must be sufficiently rigid to withstand the stress The bag tee scavenging control valve assembly may be installed on plywood attached to studs or on a...

Страница 13: ...hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certain to protect the front surface of the cabinet with a clean piece of wood between the C clamp and the...

Страница 14: ...ace steps 8 9 and then installing the bag tee scavenging control valve to the mounting bracket steps 4 7 1 6 J Hook Installation The Digital MDM Flush Cabinet Mount Installation Kit includes a J hook...

Страница 15: ...9 Figure 9 Connections Schematic some components may not be part of your installation Figure 10 Installation of Rubber Goods 1 8 Rubber Goods Installation Install the rubber goods See Figure 10...

Страница 16: ...cord at the rear of the mixer head See Figure 11 Connect the power cord to the electrical outlet WARNING For safety a Protective Earth ground connection must be made through power cord 3rd wire or Gro...

Страница 17: ...d the installer to determine that only oxygen flows into the oxygen fitting and that only nitrous oxide flows into the nitrous oxide fitting located at the rear of the mixer head If the gas pipeline t...

Страница 18: ...agent representative or dealer of Matrx by Parker shall constitute a warranty by Matrx by Parker or give rise to any liability or obligation of Matrx by Parker Subject to the next sentences Matrx by P...

Страница 19: ...9600 Vacuum gauge scavenger control valve 91525136 Swing arm assembly with mounting hardware 91505112 Pre installation kit shutoff valves 91316043 3 liter breathing bag Mounting Fasteners 91525141 Har...

Страница 20: ...ter Indexed Safety System 92305110 DISS female X DISS female Oxygen 92305111 DISS female X DISS female Nitrous oxide 92305248 DISS female X DISS female Vacuum DISS Hand I Twist 92305288 DISS Hand I Tw...

Отзывы: