CE Notice

 

 
 
The                symbol on this product indicates compliance with the Medical Device Directive of the European 
Community. A "Declaration of Conformity" in accordance with the Directive has been made and is on file. 
European Communities should contact the Authorized Representative listed below regarding any Medical Device 
Directive (MDD) inquiries. Direct inquiries may be made to the following addresses: 
 
 

Porter Regulatory Affairs: 

EU Authorized Representative: 

 
Parker Hannifin Corporation 

Parker Hannifin Ltd 

Porter Instrument  

Instrumentation Products Division

 

245 Township Line Road

  

Riverside Road, Pottington Business Park  

Hatfield, PA 19440 USA  

Barnstaple, EX31 1NP, England  

Tel: 215-723-4000 

Phone: +44 (0) 1271 313131 

 

Fax:  +44 (0) 1271 373636 

 

 

 
 
 
 

MEDICAL EQUIPMENT WITH 
RESPECT TO ELECTRIC 
SHOCK, FIRE AND 
MECHANICAL HAZARDS 
ONLYIN ACCORDANCE WITH 
UL 2601-1, ANDCAN/CSA C22.2 
NO. 601.1 

 
 

Symbol 

Description 

 
 

 

Alternating Current 

 

 

Protective Earth (ground) 

 

 

Attention, consult the accompanying documents 

 
 

 

 
 
 
 
 

Содержание Matrx Digital MDM

Страница 1: ...Ref 10550800 Rev F Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA Tel 215 723 4000...

Страница 2: ...g addresses Porter Regulatory Affairs EU Authorized Representative Parker Hannifin Corporation Parker Hannifin Ltd Porter Instrument Instrumentation Products Division 245 Township Line Road Riverside...

Страница 3: ...ility A licensed electrical person should be consulted to confirm the Protective Earth ground connection Do not use an electrical cord with nicks cuts or other damage Do not use the unit with an exten...

Страница 4: ...enging Control Valve Installation 6 1 6 J Hook Installation 8 1 7 Hose Connection 8 1 8 Rubber Goods Installation 9 1 9 Installation of Power Cord and Protective Earth ground Connection 10 Section 2 S...

Страница 5: ...ns 5 Figure 4 Screw and Tee Nut Adjustment 5 Figure 5 Digital MDM Mixer Head Installation 6 Figure 6 Flush Mounting the Bag Tee and Scavenging Control Valve 6 Figure 7 Surface Mounting the Bag Tee and...

Страница 6: ...tact Information Unit Serial Number Date Purchased Dealer Name Dealer Phone Number Notes Contact Information Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA...

Страница 7: ...he fittings and tubing incorporate unique tubing sizes for each gas an important safety feature to prevent accidental crossing of lines during installation or maintenance The following basic requireme...

Страница 8: ...information on the NFPA 99C code requirements consult your local Matrx by Parker representative or your local dental dealer U S Gas Color Coding All products in this manual are manufactured according...

Страница 9: ...A guideline No 99C see the Introduction for details All equipment must be installed in accordance with these guidelines as well as all local codes and ordinances Matrx by Parker makes no warranty for...

Страница 10: ...tion 1 Install the necessary gas plumbing from the central gas supply 2 If separate gas outlet stations are to be installed install the necessary gas plumbing See the Outlet Stations Installation and...

Страница 11: ...4 mounting holes See Figure 4 6 While supporting the front of the cabinet tap in the tee nuts with a hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certa...

Страница 12: ...installed The attachment to the cabinet surface must be sufficiently rigid to withstand the stress The bag tee scavenging control valve assembly may be installed on plywood attached to studs or on a...

Страница 13: ...hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certain to protect the front surface of the cabinet with a clean piece of wood between the C clamp and the...

Страница 14: ...ace steps 8 9 and then installing the bag tee scavenging control valve to the mounting bracket steps 4 7 1 6 J Hook Installation The Digital MDM Flush Cabinet Mount Installation Kit includes a J hook...

Страница 15: ...9 Figure 9 Connections Schematic some components may not be part of your installation Figure 10 Installation of Rubber Goods 1 8 Rubber Goods Installation Install the rubber goods See Figure 10...

Страница 16: ...cord at the rear of the mixer head See Figure 11 Connect the power cord to the electrical outlet WARNING For safety a Protective Earth ground connection must be made through power cord 3rd wire or Gro...

Страница 17: ...d the installer to determine that only oxygen flows into the oxygen fitting and that only nitrous oxide flows into the nitrous oxide fitting located at the rear of the mixer head If the gas pipeline t...

Страница 18: ...agent representative or dealer of Matrx by Parker shall constitute a warranty by Matrx by Parker or give rise to any liability or obligation of Matrx by Parker Subject to the next sentences Matrx by P...

Страница 19: ...9600 Vacuum gauge scavenger control valve 91525136 Swing arm assembly with mounting hardware 91505112 Pre installation kit shutoff valves 91316043 3 liter breathing bag Mounting Fasteners 91525141 Har...

Страница 20: ...ter Indexed Safety System 92305110 DISS female X DISS female Oxygen 92305111 DISS female X DISS female Nitrous oxide 92305248 DISS female X DISS female Vacuum DISS Hand I Twist 92305288 DISS Hand I Tw...