8
3.
Drill the (2) 3/8” mounting holes.
4. Slide a screw cap onto each of the (4) short screws,
as in Figure 4.
5. If the practitioner wants the hoses to attach from
below, orient the hose barbs (they turn) parallel to
the mounting plate. If the practitioner wants the
hoses to attach from behind, orient the hose barbs
perpendicular to the mounting plate.
6. Align the mounting plate of the bag tee/scavenging
control valve assembly to the front of the mounting
bracket. Secure with the (4) screws.
7. Tighten the (4) screws and snap the screw caps
over the screw heads.
8. Align the back of the mounting bracket with the (2)
holes on the cabinet surface, and secure with the
(2) 2-
1/4” bolts.
9. From the other side of the cabinet, install a flat
washer, a lock washer, and a nut on each bolt.
Tighten the nuts.
10. Connect the gas hoses to the hose barbs. See
Section 1.7.
NOTE
If desired, the installer may reverse the order of
installation, first installing the mounting bracket to
the cabinet surface (steps 8
– 9) and then installing
the bag tee/scavenging control valve to the
mounting bracket (steps 4
– 7).
1.6
J Hook Installation
The Digital MDM® Flush Cabinet Mount Installation Kit
includes a J hook for rubber goods storage. Mount the J
hook in a convenient location; consult with the
practitioner.
1.7
Hose Connections
WARNING
The mixed gas outlet from the rear of the mixer
head must be connected to a single bag tee only.
For gas compatibility, all hoses must be supplied by
Matrx by Parker. These hoses are braided
to prevent occlusion when bent.
All hoses should be as short as possible and
installed in a manner that prevents kinking,
crushing, and excessive movement. Properly
installed, short hoses minimize back pressure on
the system.
All hoses must be fully seated on the hose barbs.
This prevents gas leakage, which would result in
the mixer’s incorrectly displaying the volume of gas
being delivered to the patient.
Following the schematic in Figure 9, install all hoses at
the proper connections. Be aware that your installation
may not include all the components shown in Figure 9.
Figure 8. Mounting the Bag Tee and Scavenging Control Valve Inside the Cabinet