background image

 
 
 

3. 

Drill the (2) 3/8” mounting holes. 

 
4.  Slide a screw cap onto each of the (4) short screws, 

as in Figure 4. 

 
5.  If the practitioner wants the hoses to attach from 

below, orient the hose barbs (they turn) parallel to 
the mounting plate. If the practitioner wants the 
hoses to attach from behind, orient the hose barbs 
perpendicular to the mounting plate. 

 
6.  Align the mounting plate of the bag tee/scavenging 

control valve assembly to the front of the mounting 
bracket. Secure with the (4) screws. 

 
7.  Tighten the (4) screws and snap the screw caps 

over the screw heads. 

 
8.  Align the back of the mounting bracket with the (2) 

holes on the cabinet surface, and secure with the 
(2) 2-

1/4” bolts. 

 
9.  From the other side of the cabinet, install a flat 

washer, a lock washer, and a nut on each bolt. 
Tighten the nuts. 

 
10.  Connect the gas hoses to the hose barbs. See 

Section 1.7. 

 

NOTE 

If desired, the installer may reverse the order of 
installation, first installing the mounting bracket to 
the cabinet surface (steps 8 

– 9) and then installing 

the bag tee/scavenging control valve to the 
mounting bracket (steps 4 

– 7).  

 
1.6 

J Hook Installation 

The Digital MDM® Flush Cabinet Mount Installation Kit 
includes a J hook for rubber goods storage. Mount the J 
hook in a convenient location; consult with the 
practitioner. 
 

1.7 

Hose Connections 

 

WARNING 

The mixed gas outlet from the rear of the mixer 
head must be connected to a single bag tee only. 
 
For gas compatibility, all hoses must be supplied by 
Matrx by Parker. These hoses are braided  
to prevent occlusion when bent. 
 
All hoses should be as short as possible and 
installed in a manner that prevents kinking, 
crushing, and excessive movement. Properly 
installed, short hoses minimize back pressure on 
the system. 
 
All hoses must be fully seated on the hose barbs. 
This prevents gas leakage, which would result in 
the mixer’s incorrectly displaying the volume of gas 
being delivered to the patient. 

 
Following the schematic in Figure 9, install all hoses at 
the proper connections. Be aware that your installation 
may not include all the components shown in Figure 9. 
 
 

 

 

Figure 8.  Mounting the Bag Tee and Scavenging Control Valve Inside the Cabinet 

Содержание Matrx Digital MDM

Страница 1: ...Ref 10550800 Rev F Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA Tel 215 723 4000...

Страница 2: ...g addresses Porter Regulatory Affairs EU Authorized Representative Parker Hannifin Corporation Parker Hannifin Ltd Porter Instrument Instrumentation Products Division 245 Township Line Road Riverside...

Страница 3: ...ility A licensed electrical person should be consulted to confirm the Protective Earth ground connection Do not use an electrical cord with nicks cuts or other damage Do not use the unit with an exten...

Страница 4: ...enging Control Valve Installation 6 1 6 J Hook Installation 8 1 7 Hose Connection 8 1 8 Rubber Goods Installation 9 1 9 Installation of Power Cord and Protective Earth ground Connection 10 Section 2 S...

Страница 5: ...ns 5 Figure 4 Screw and Tee Nut Adjustment 5 Figure 5 Digital MDM Mixer Head Installation 6 Figure 6 Flush Mounting the Bag Tee and Scavenging Control Valve 6 Figure 7 Surface Mounting the Bag Tee and...

Страница 6: ...tact Information Unit Serial Number Date Purchased Dealer Name Dealer Phone Number Notes Contact Information Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA...

Страница 7: ...he fittings and tubing incorporate unique tubing sizes for each gas an important safety feature to prevent accidental crossing of lines during installation or maintenance The following basic requireme...

Страница 8: ...information on the NFPA 99C code requirements consult your local Matrx by Parker representative or your local dental dealer U S Gas Color Coding All products in this manual are manufactured according...

Страница 9: ...A guideline No 99C see the Introduction for details All equipment must be installed in accordance with these guidelines as well as all local codes and ordinances Matrx by Parker makes no warranty for...

Страница 10: ...tion 1 Install the necessary gas plumbing from the central gas supply 2 If separate gas outlet stations are to be installed install the necessary gas plumbing See the Outlet Stations Installation and...

Страница 11: ...4 mounting holes See Figure 4 6 While supporting the front of the cabinet tap in the tee nuts with a hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certa...

Страница 12: ...installed The attachment to the cabinet surface must be sufficiently rigid to withstand the stress The bag tee scavenging control valve assembly may be installed on plywood attached to studs or on a...

Страница 13: ...hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certain to protect the front surface of the cabinet with a clean piece of wood between the C clamp and the...

Страница 14: ...ace steps 8 9 and then installing the bag tee scavenging control valve to the mounting bracket steps 4 7 1 6 J Hook Installation The Digital MDM Flush Cabinet Mount Installation Kit includes a J hook...

Страница 15: ...9 Figure 9 Connections Schematic some components may not be part of your installation Figure 10 Installation of Rubber Goods 1 8 Rubber Goods Installation Install the rubber goods See Figure 10...

Страница 16: ...cord at the rear of the mixer head See Figure 11 Connect the power cord to the electrical outlet WARNING For safety a Protective Earth ground connection must be made through power cord 3rd wire or Gro...

Страница 17: ...d the installer to determine that only oxygen flows into the oxygen fitting and that only nitrous oxide flows into the nitrous oxide fitting located at the rear of the mixer head If the gas pipeline t...

Страница 18: ...agent representative or dealer of Matrx by Parker shall constitute a warranty by Matrx by Parker or give rise to any liability or obligation of Matrx by Parker Subject to the next sentences Matrx by P...

Страница 19: ...9600 Vacuum gauge scavenger control valve 91525136 Swing arm assembly with mounting hardware 91505112 Pre installation kit shutoff valves 91316043 3 liter breathing bag Mounting Fasteners 91525141 Har...

Страница 20: ...ter Indexed Safety System 92305110 DISS female X DISS female Oxygen 92305111 DISS female X DISS female Nitrous oxide 92305248 DISS female X DISS female Vacuum DISS Hand I Twist 92305288 DISS Hand I Tw...

Отзывы: