SAFETY SUMMARY 

 

 WARNINGS 

The Matrx by Parker DIGITAL MDM® mixer is a specialized medical device and should be operated ONLY by a 
professional trained in the use of nitrous oxide. Nitrous oxide and oxygen mixtures should only be administered by 
or under the supervision of a medical professional. 

 

The Matrx by Parker DIGITAL MDM mixer is designed for use with an oxygen supply assuring 100% medical 
oxygen concentration. The use of other oxygen sources (i.e., oxygen concentrators) could result in an unsafe 
condition to the patient. 

 

Although this analgesia machine is designed to prevent the interchanging of the two gases, it is the responsibility 
of the user to determine that only oxygen flows into the oxygen connection and only nitrous oxide flows into the 
nitrous oxide connection located on the back of the analgesia head. 

 

ELECTRICAL CONNECTIONS & GROUNDING INSTRUCTIONS 
The unit must be grounded.  If it should malfunction or breakdown, grounding provides a path of 
least resistance for electrical current to reduce the risk of electric shock.   

 

A known Protective Earth (ground) connection must be made via one of the following methods:  
1)  The third wire on the power cord when plugged into an appropriate outlet that is properly 

installed and grounded in accordance with all local codes or ordinances.  

2)  The external ground lug on the rear of the device.   

 

A Protective Earth (ground) connection is a connection attached to the facility grounding point 
outside the facility.  A licensed electrical person should be consulted to confirm the Protective 
Earth (ground) connection. 

 

 

Do not use an electrical cord with nicks, cuts, or other damage. Do not use the unit with an 
extension cord as it will be subject to electromagnetic interference (EMI). 

 

BAG TEE REQUIREMENTS 
A bag tee containing an air inlet valve is required for use with this unit. The bag tee is factory 
mounted on the unit or provided separately for remote mounting. Matrx by Parker Scavenging 
Nasal Inhalers provide the patient connection. 

 

CAUTIONS               

FACTORY SETTINGS 
Do not change any internal factory set adjustments. The Matrx DIGITAL MDM mixer is adjusted and calibrated for 
proper operation prior to shipment. Any change could alter the specified operation and accuracy. 

 

FACILITY PIPING 
The user MUST perform tests independent of the dealer or contractor to verify that all pipelines are connected 
correctly. If you are unfamiliar with piping systems, order Matrx manual 10311700. The user has the ultimate 
responsibility for properly connected pipelines essential to patient safety. 

 

DIAMETER INDEXED SAFETY SYSTEM 
Gas connections in the DIGITAL MDM mixer are mechanically indexed using the Diameter Indexed Safety 
System (DISS). This feature helps to prevent the crossing of pipelines. Under no circumstances should these 
connections be altered. 

 

DISCONNECT POWER 
There are no user-serviceable parts inside this unit. A shock hazard exists if power is not 
disconnected prior to any internal servicing or maintenance. Always disconnect the power 
source before inspection or repair. 

 

ALARM TEST 
The user should test the alarm system for proper operation prior to each clinical use. 

Содержание Matrx Digital MDM

Страница 1: ...Ref 10550800 Rev F Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA Tel 215 723 4000...

Страница 2: ...g addresses Porter Regulatory Affairs EU Authorized Representative Parker Hannifin Corporation Parker Hannifin Ltd Porter Instrument Instrumentation Products Division 245 Township Line Road Riverside...

Страница 3: ...ility A licensed electrical person should be consulted to confirm the Protective Earth ground connection Do not use an electrical cord with nicks cuts or other damage Do not use the unit with an exten...

Страница 4: ...enging Control Valve Installation 6 1 6 J Hook Installation 8 1 7 Hose Connection 8 1 8 Rubber Goods Installation 9 1 9 Installation of Power Cord and Protective Earth ground Connection 10 Section 2 S...

Страница 5: ...ns 5 Figure 4 Screw and Tee Nut Adjustment 5 Figure 5 Digital MDM Mixer Head Installation 6 Figure 6 Flush Mounting the Bag Tee and Scavenging Control Valve 6 Figure 7 Surface Mounting the Bag Tee and...

Страница 6: ...tact Information Unit Serial Number Date Purchased Dealer Name Dealer Phone Number Notes Contact Information Parker Hannifin Corporation Porter Instrument 245 Township Line Road Hatfield PA 19440 USA...

Страница 7: ...he fittings and tubing incorporate unique tubing sizes for each gas an important safety feature to prevent accidental crossing of lines during installation or maintenance The following basic requireme...

Страница 8: ...information on the NFPA 99C code requirements consult your local Matrx by Parker representative or your local dental dealer U S Gas Color Coding All products in this manual are manufactured according...

Страница 9: ...A guideline No 99C see the Introduction for details All equipment must be installed in accordance with these guidelines as well as all local codes and ordinances Matrx by Parker makes no warranty for...

Страница 10: ...tion 1 Install the necessary gas plumbing from the central gas supply 2 If separate gas outlet stations are to be installed install the necessary gas plumbing See the Outlet Stations Installation and...

Страница 11: ...4 mounting holes See Figure 4 6 While supporting the front of the cabinet tap in the tee nuts with a hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certa...

Страница 12: ...installed The attachment to the cabinet surface must be sufficiently rigid to withstand the stress The bag tee scavenging control valve assembly may be installed on plywood attached to studs or on a...

Страница 13: ...hammer Alternately a C clamp can be used to squeeze the tee nuts into the cabinet surface Be certain to protect the front surface of the cabinet with a clean piece of wood between the C clamp and the...

Страница 14: ...ace steps 8 9 and then installing the bag tee scavenging control valve to the mounting bracket steps 4 7 1 6 J Hook Installation The Digital MDM Flush Cabinet Mount Installation Kit includes a J hook...

Страница 15: ...9 Figure 9 Connections Schematic some components may not be part of your installation Figure 10 Installation of Rubber Goods 1 8 Rubber Goods Installation Install the rubber goods See Figure 10...

Страница 16: ...cord at the rear of the mixer head See Figure 11 Connect the power cord to the electrical outlet WARNING For safety a Protective Earth ground connection must be made through power cord 3rd wire or Gro...

Страница 17: ...d the installer to determine that only oxygen flows into the oxygen fitting and that only nitrous oxide flows into the nitrous oxide fitting located at the rear of the mixer head If the gas pipeline t...

Страница 18: ...agent representative or dealer of Matrx by Parker shall constitute a warranty by Matrx by Parker or give rise to any liability or obligation of Matrx by Parker Subject to the next sentences Matrx by P...

Страница 19: ...9600 Vacuum gauge scavenger control valve 91525136 Swing arm assembly with mounting hardware 91505112 Pre installation kit shutoff valves 91316043 3 liter breathing bag Mounting Fasteners 91525141 Har...

Страница 20: ...ter Indexed Safety System 92305110 DISS female X DISS female Oxygen 92305111 DISS female X DISS female Nitrous oxide 92305248 DISS female X DISS female Vacuum DISS Hand I Twist 92305288 DISS Hand I Tw...

Отзывы: