Bulletin TI-N2-1100K
Installation, Operation and Maintenance Manual
Quick Start Summary
Parker Balston Model HPN2-1100, UHPN2-1100
www.balstonlabgasgenerators.com
2
1-800-343-4048
Quick Start
Summary
Parker Balston HPN2-1100 and
UHPN2-1100 Ultra Pure Nitrogen
Generators
Installation, Operation, and Maintenance Manual
This Quick Start Summary is designed to get the Parker Balston Nitrogen Generator in-
stalled and operational quickly. This is not intended to replace the full instructions. Please
be sure to read this entire manual.
Ambient Temperature
(page 4):
60°F - 100°F (16°C - 38°C)
Electrical (page 4):
User selectable: 120 VAC or 230 VAC, 50/60 Hz
Compressed Ai
r
(page 4):
Inlet air line: 1/4" or larger
Minimum Compressor capacity: 1.5 scfm (45 lpm)
Air quality: instrument quality (see "Accessories", page 8)
Chlorinated hydrocarbon compounds and chlorofluorocarbons (or freons) will permanently
damage the hydrocarbon catalyst module. Extreme care should be taken when specifying
an air supply for the generator to ensure that these compounds are not present in the air
supply or stored near the compressor intake.
Recommended Accessories:
See technical information bulletin.
Outlet Piping:
Stainless steel or refrigeration grade copper, outlet pressure regulator, flow
controller.
Installation
(see technical information
bulletin for details)
Bulletin TI-N2-1100K