background image

Bulletin TI-N2-1100K 

Installation, Operation and Maintenance Manual

Quick Start Summary 

Parker Balston Model HPN2-1100, UHPN2-1100

www.balstonlabgasgenerators.com

5

1-800-343-4048

Hydrocarbon Removal

(Model UHPN2-1100 only)

Oxygen Removal

Final Filtration

In addition to the coalescing prefilter, particulate filters are incorporated into the design of the 

Nitrogen Generator to protect internal components from particulate contamination.  The Model 

UHPN2-1100 has two Grade BX Balston particulate filters, and the Model HPN2-1100 has one 

Grade BX particulate filter as shown on the Flow Schematics (Figures 2 and 3).
The catalyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst 

and assembled with a cartridge heater.  This catalyst module oxidizes hydrocarbons in the 

compressed air supply to water vapor and carbon dioxide, which are removed in subsequent 

purification steps.

The next stage in the purification process of the Parker Balston Nitrogen Generator is the 

removal of oxygen, water vapor, and carbon dioxide from the compressed air supply.  In this 

stage, pressure swing adsorption (PSA) technology is utilized to reduce the concentrations of 

these compounds to <1 ppm (each). 

The final filter on the Parker Balston Nitrogen Generator is a Balston Grade GS membrane 

filter which removes particulate contamination to 0.01 micron (absolute).

Figure 2

 - 

HPN2-1100 Flow Schematic

Figure 3

 - 

UHPN2-1100 Flow Schematic

Содержание Ultra Pure HPN2-1100

Страница 1: ...ntenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 1 1 800 343 4048 Parker Balston Ultra Pure Nitrogen Generators Installation Operation and...

Страница 2: ...16 C 38 C Electrical page 4 User selectable 120 VAC or 230 VAC 50 60 Hz Compressed Air page 4 Inlet air line 1 4 or larger Minimum Compressor capacity 1 5 scfm 45 lpm Air quality instrument quality se...

Страница 3: ...pply valve and set the inlet pressure regulator see illustration on page 2 Turn the power switch upward to the On position The System Check light should blink every 12 seconds 3 The nitrogen generator...

Страница 4: ...i trogen Generators are designed to provide high purity nitrogen gas for analytical instrumenta tion and other laboratory applications The HP UHP Nitrogen Generators are certified to the electrical re...

Страница 5: ...alyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst and assembled with a cartridge heater This catalyst module oxidizes hydrocarbons in the compressed air supply...

Страница 6: ...catalyst module heater Model UHPN2 1100 only or the electrical connections to it This indicator light is inoperative in the Model HPN2 1100 generator Inlet Pressure Display The inlet pressure display...

Страница 7: ...to the compressor providing air to the generator The hydrocarbon catalyst module can also be contaminated by high concentrations of lead sulfur or phosphorous compounds heavy metals and long chain pol...

Страница 8: ...are incorporated into the downstream equipment being used there is no need to install them separately These accessories must be appropriate for high purity applications Pressure Regulator downstream...

Страница 9: ...ween 60 psig and 125 psig 4 1 barg to 8 6 barg During the 5 hours of warm up there will be no flow of nitrogen from the generator Figure 5 Manual Shut Off Valve Indicating Oxygen Trap If visible verif...

Страница 10: ...the power returns If the power supply is interrupted for more than 1 hour the initial 5 hour start up cycle will be repeated when the power returns The flow of nitrogen from the Parker Balston Nitrog...

Страница 11: ...been assembled into a maintenance kit MK7692 Maintenance Kit for Model HPN2 1100 and MK7694 Maintenance Kit for Model UHPN2 1100 If necessary the Nitrogen Generator may be wiped clean with a dry cloth...

Страница 12: ...ion Operation and Maintenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 12 1 800 343 4048 Figure 6 Maintenance Items PARTICULATE FILTER P N 7...

Страница 13: ...egrity of the product use only fuses of the size and type detailed in the specifications section of this bulletin This equipment has main fuses in both neutral and phase lines Use care when servicing...

Страница 14: ...1100 UHPN2 1100 System Specifications Table 1 Nitrogen Outlet Pressure and Flow Specifications Inlet Air Pressure Maximum N2 Flow Maximum N2 Pressure 125 psig 8 6 barg 1 1 lpm 85 psig 5 9 barg 110 ps...

Страница 15: ...ations can result in unconsciousness without any warning symptoms such as dizziness fatigue etc Cautions WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin g...

Страница 16: ...Symbols 3 CELL UHP Printed in U S A Bulletin TI N2 1100K 03 2017 2003 2017 Parker Hannifin Corporation Parker Manufacturing UK Ltd Gas Separation Filtration Division EMEA Dukesway Team Valley Trading...

Отзывы: