Bulletin TI-N2-1100K 

Installation, Operation and Maintenance Manual

Quick Start Summary 

Parker Balston Model HPN2-1100, UHPN2-1100

www.balstonlabgasgenerators.com

10

1-800-343-4048

After the initial start-up cycle (5 hours) has been completed, the “System Check” LED will go 

off and the “System Ready” LED will illuminate.  Open the external outlet valve and the gen-

erator will supply ultra high purity nitrogen to the application.  Check all downstream piping 

connections for leaks using a bubble-type leak detecting solution.  

If there is a power or pressure interruption, the “System Check” LED will go on and the sys-

tem may go back into a start-up mode.  (See “Power Interruptions” and “Pressure Interrup-

tions” sections below).  If the “System Ready” LED indicator goes off and the “Service System” 

LED indicator is illuminated or blinking, the operator should check the hydrocarbon catalyst 

module (Model UHPN2-1100 only).  It may need to be replaced as part of the normal mainte-

nance of the hydrocarbon catalyst module, or it may have a loose electrical connection (see 

the Troubleshooting Section at the end of this bulletin).

If the compressed air pressure falls below 60 psig (4.1 barg), an internal pressure transducer 

will activate an audible alarm and the yellow “System Check” light will be illuminated to 

attract operator attention.  If the pressure remains below 60 psig (4.1 barg) for less than 6 

minutes (time required for purge and fill of PSA towers), the system will continue to operate 

at rated flow and purity.  If the inlet air pressure drops below 60 psig (4.1 barg) for more than                  

6 minutes, the internal outlet valve will close and the flow of nitrogen will cease.  After this 

low pressure condition is corrected, the outlet valve will open and the nitrogen flow will start 

again immediately. 

If the electrical power is interrupted for less than 30 minutes, the system will deliver nitro-

gen at rated flow and purity when the power returns.  If the power is interrupted for between           

30 and 60 minutes, the generator will go into an abbreviated 1 hour start-up cycle when the 

power returns.  If the power supply is interrupted for more than 1 hour, the initial 5 hour start-

up cycle will be repeated when the power returns.

The flow of nitrogen from the Parker Balston Nitrogen Generator must be controlled using 

a metering valve or flow controller downstream from the unit.  The flow control device may 

be integral to the equipment being used downstream, or it may be an independent controller 

(see Installation Section) purchased and installed by the customer.  In either case, a device to 

measure the flow should also be installed (either with the flow controller or downstream from 

the application).  If the demand of ultra high purity nitrogen exceeds the delivery capacity of 

the generator, the purity and pressure of the emergent nitrogen stream will not meet specifi-

cations.  

The outlet pressure and flow from the generator must be adjusted according to the operating 

parameters listed below.  If the outlet flow exceeds the rated flow for a given inlet pressure, 

the nitrogen purity will not meet specifications.

Pressure Interruptions

Operation

Power Interruptions

 Minimum 

Maximum 

Maximum 

Receiver

 

Inlet Pressure 

Outlet Flow 

Outlet Pressure 

Pressure*

  120 psig   8.3 barg 

1.1 lpm 

85 psig   5.9 barg 

90-115 psig   6.2-7.9 barg

  110 psig   7.6 barg 

1.0 lpm 

75 psig   5.2 barg 

80-100 psig   5.5-6.9 barg

  100 psig   6.9 barg 

0.9 lpm 

65 psig   4.5 barg 

70-90 psig   4.8-6.2 barg

  90 psig   6.2 barg 

0.8 lpm 

60 psig   4.1 barg 

65-80 psig   4.5-5.5 barg

  80 psig   5.5 barg 

0.7 lpm 

50 psig   3.4 barg 

56-73 psig   3.9-5.0 barg

  70 psig   4.8 barg 

0.6 lpm 

45 psig   3.1 barg 

50-64 psig   3.4-4.4 barg

  60 psig   4.1 barg 

0.5 lpm 

35 psig   2.4 barg 

40-52 psig   2.8-3.6 barg

Outlet Flow Operating 

Parameters

If the generator will be shut down for longer than a few hours, turn off the power to the gen-

erator and close the (customer installed) inlet air valve.   Allow the generator to purge for 10 

minutes.  Open the lower service panel and close the 2-way manual valve upstream from the 

PSA towers (see Figure 5).  Finally, close the external manual outlet valve.  

If the generator is 

not shut down according to this procedure, the PSA towers may be poisoned and the war-

ranty will be void.

Shutdown

Flow Rate Adjustment

* Note:  During the warmup, receiver pressure gauge reading will be zero.

Содержание Ultra Pure HPN2-1100

Страница 1: ...ntenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 1 1 800 343 4048 Parker Balston Ultra Pure Nitrogen Generators Installation Operation and...

Страница 2: ...16 C 38 C Electrical page 4 User selectable 120 VAC or 230 VAC 50 60 Hz Compressed Air page 4 Inlet air line 1 4 or larger Minimum Compressor capacity 1 5 scfm 45 lpm Air quality instrument quality se...

Страница 3: ...pply valve and set the inlet pressure regulator see illustration on page 2 Turn the power switch upward to the On position The System Check light should blink every 12 seconds 3 The nitrogen generator...

Страница 4: ...i trogen Generators are designed to provide high purity nitrogen gas for analytical instrumenta tion and other laboratory applications The HP UHP Nitrogen Generators are certified to the electrical re...

Страница 5: ...alyst module used for hydrocarbon removal is a stainless steel vessel filled with catalyst and assembled with a cartridge heater This catalyst module oxidizes hydrocarbons in the compressed air supply...

Страница 6: ...catalyst module heater Model UHPN2 1100 only or the electrical connections to it This indicator light is inoperative in the Model HPN2 1100 generator Inlet Pressure Display The inlet pressure display...

Страница 7: ...to the compressor providing air to the generator The hydrocarbon catalyst module can also be contaminated by high concentrations of lead sulfur or phosphorous compounds heavy metals and long chain pol...

Страница 8: ...are incorporated into the downstream equipment being used there is no need to install them separately These accessories must be appropriate for high purity applications Pressure Regulator downstream...

Страница 9: ...ween 60 psig and 125 psig 4 1 barg to 8 6 barg During the 5 hours of warm up there will be no flow of nitrogen from the generator Figure 5 Manual Shut Off Valve Indicating Oxygen Trap If visible verif...

Страница 10: ...the power returns If the power supply is interrupted for more than 1 hour the initial 5 hour start up cycle will be repeated when the power returns The flow of nitrogen from the Parker Balston Nitrog...

Страница 11: ...been assembled into a maintenance kit MK7692 Maintenance Kit for Model HPN2 1100 and MK7694 Maintenance Kit for Model UHPN2 1100 If necessary the Nitrogen Generator may be wiped clean with a dry cloth...

Страница 12: ...ion Operation and Maintenance Manual Quick Start Summary Parker Balston Model HPN2 1100 UHPN2 1100 www balstonlabgasgenerators com 12 1 800 343 4048 Figure 6 Maintenance Items PARTICULATE FILTER P N 7...

Страница 13: ...egrity of the product use only fuses of the size and type detailed in the specifications section of this bulletin This equipment has main fuses in both neutral and phase lines Use care when servicing...

Страница 14: ...1100 UHPN2 1100 System Specifications Table 1 Nitrogen Outlet Pressure and Flow Specifications Inlet Air Pressure Maximum N2 Flow Maximum N2 Pressure 125 psig 8 6 barg 1 1 lpm 85 psig 5 9 barg 110 ps...

Страница 15: ...ations can result in unconsciousness without any warning symptoms such as dizziness fatigue etc Cautions WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin g...

Страница 16: ...Symbols 3 CELL UHP Printed in U S A Bulletin TI N2 1100K 03 2017 2003 2017 Parker Hannifin Corporation Parker Manufacturing UK Ltd Gas Separation Filtration Division EMEA Dukesway Team Valley Trading...

Отзывы: