Bulletin TI-N2-1100K
Installation, Operation and Maintenance Manual
Quick Start Summary
Parker Balston Model HPN2-1100, UHPN2-1100
www.balstonlabgasgenerators.com
14
1-800-343-4048
IEC 1010 Installation
Category II
IEC 1010 Pollution
Degree 2
CSA Certification Standard
CAN/CSA 22.2 No. 1010-92
IEC Safety Standard
IEC 1010-1:1990+ A1 1992+ A2:1995/ENG1010-1:1993
UL Safety Standard
UL3101-1, First Edition
UL Listed to U.S. and Canadian Safety Standards
Nitrogen Purity
99.9999% 99.9999%
O
2
, H
2
O, CO
2,
CO Concentration
< 1.0 ppm
< 1.0 ppm
Hydrocarbon Concentration
N/A
< 0.1 ppm
Ar Concentration
0.9%
0.9%
Inlet Connection
1/4" NPT (female)
1/4" NPT (female)
Outlet Connection
1/8" NPT (female)
1/8" NPT (female)
Compressed Air Consumption
1.5 SCFM (42 lpm)
1.5 SCFM (42 lpm)
Min/Max Inlet Air Pressure
(4)
60 psig/125 psig (4.1 barg/8.6 barg)
60 psig/125 psig (4.1 barg/8.6 barg)
Recommended Inlet Air Temperature
78°F (26°C)
78°F (26°C)
Altitude
2000 m
2000 m
Max Inlet Air Dewpoint
60°F (15°C)
60°F (15°C)
Ambient Operating Temperature
60°F to 100°F (16°C to 38°C)
60°F to 100°F (16°C to 38°C)
Electrical Requirements
120VAC / 230VAC, 50/60 Hz, 25 watts
120VAC / 230VAC, 50/60 Hz, 575 watts
Dimensions
12"w X 16"d X 35"h
12"w X 16"d X 35"h
(31cm X 41cm X 89cm)
(31cm X 41cm X 89cm)
Shipping Weight
110 lbs. (50 kg)
110 lbs. (50 kg)
Fuse Type (120 VAC Model)
1 amp, Type T, 250 V
5 amp, Type T, 250 V
(230 VAC Model)
1 amp, Type T, 250 V
3.15 amp, Type T, 250 V
(100 VAC Model)
1 amp, Type T, 250 V
4 amp, Type T, 250 V
HPN2-1100
UHPN2-1100
System Specifications
Table 1
Nitrogen Outlet Pressure and Flow Specifications
Inlet Air Pressure
Maximum N
2
Flow
Maximum N
2
Pressure
125 psig
8.6 barg
1.1 lpm
85 psig
5.9 barg
110 psig
7.6 barg
1.0 lpm
75 psig
5.2 barg
100 psig
6.9 barg
0.9 lpm
65 psig
4.5 barg
90 psig
6.2 barg
0.8 lpm
60 psig
4.1 barg
80 psig
5.5 barg
0.7 lpm
50 psig
3.4 barg
70 psig
4.8 barg
0.6 lpm
45 psig
3.1 barg
60 psig
4.1 barg
0.5 lpm
35 psig
2.4 barg
Notes
1
The Parker Balston Nitrogen Generator
was designed to meet the specifications
listed when operated in a manner consis-
tent with the recommendations presented
in this manual. If these operating specifi
-
cations are not met, then inferior product
performance and shortened product life
may result.
2
The nitrogen generator should be run
continuously to optimize the performance
and reliability of the unit.
4
The Parker Balston Nitrogen Generator
installation should be thoroughly checked
for leaks after any service activity has been
performed on the system in particular or its
installation on the whole. Any leak could
significantly reduce the purity of the outlet
nitrogen stream.
5
The Parker Balston Nitrogen Generator will
not operate with an inlet air pressure less
than 60 psig (4.1 barg). A pressure regulator
must be installed upstream from the gen-
erator to maintain an air supply pressure
of 60 psig to 125 psig (4.1 barg to 8.6
barg).
Optional Accessories
P/N
Indicating Oxygen Trap
72092
Prefilter, if needed
2002N-1B1-DX
Pressure Regulator
(upstream) 72-130-V883
Pressure Regulator
(downstream)
W-425-4032-000
Surge Supressor
75043 (North America) 75044 (Europe)
Hydrocarbon Scrubber
76080