background image

19

1

 Press the hook of the filter case on the side of the pump and 

take out the case.

2

 Replace the filter. 

Reset the filter operating time after re-

placing the filter.

 

„

How to clean the air cells

1

 Remove the couplings (upper and lower) from the pump and discharge air from the air cells.

2

 Remove the top cover.

3

 Clean the air cells using the same method as the covers. (See page 18)

4

 Dry the air cells naturally and always connect the couplings (upper and lower) to the pump properly. (See page 6)

5

 Close the cover zippers properly.

CAUTION

After maintenance, check that the connector for the blowing tubes are not disconnected. Air 

may not be supplied.

Note

!

 

z

When maintaining the air cells, do not separate each air cell from the mattress.

 

z

If the mattress is used for a long period of time, the urethane film of the air cells may be discolored but there 

will be no problem with performance.

 

„

 How to clean the pump and operation panel

1

 Use soapy water (commercially available soap, such as neutral detergent diluted with water) to wipe and clean the pump and 

operation panel.

2

 Use disinfectant to clean and disinfect the pump and operation panel. For disinfectant, use the following chemicals and observe 

the specified concentration. Also, use those disinfectants according to the handling method of each disinfectant.

 

z

76.9% to 81.4% Ethanol for disinfection

 

z

0.05% to 0.2%

Benzethonium chloride

 

z

70%

Isopropanol

 

z

0.05%

Chlorhexidine gluconate

 

z

0.05% to 0.2%

Benzalkonium chloride

 

z

0.02% to 1.0%

Sodium hypochlorite

 

„

 

Replacing the filter

Regularly inspect and replace the filter. For the details on the dedicated filter, contact Paramount Bed (see page 31) or your 

distributor.

CAUTION

 

z

Regularly inspect the filter. If the filter is clogged, it 

may not work properly.

 

z

If a filter other than specified is attached or no filter is 

attached, insects or dust may enter the inside of the 

pump, possibly causing malfunction. This also may 

cause failure.

Take out

Pump side

Filter

Filter case

Press hook

 

„

 

How to clean the glide sheet and top cushion

1

 Open the zippers and then remove the top cover from the bottom cover.

2

 Remove the belts of the air cell from the rubber loops of the bottom cover at the head end two corners, and then remove the 

belts of the air cell from the loops of the glide sheet. Remove the strings from the loops at the foot end two corners. Untie the 

strings from the loops at the two corners and then remove both the glide sheet and top cushion together.

3

 Use soapy water (commercially available soap, such as neutral detergent diluted with water) to soak the glide sheet and top 

cushion to wash them together without separating them.

4

 Rinse with water until there is no residual detergent and wring out gently without twisting. Dry completely in a shaded area.

[How to attach the glide sheet]

1

 Lay the glide sheet on the air cells so that the tag is facing up.

2

 Pass the belts of the air cell through the loops of the glide 

sheet at the head end two corners, and then attach the rub-

ber loops of the bottom cover to the belts of the air cell. 

3

 Pass the strings at the foot end two corners of the glide sheet 

through the loops from the outer side to the inner side of the 

bottom cover and then tie securely.

Note

!

 

z

When attaching the glide sheet, take care that 

the urethane sheet inside does not shift toward 

one side.

 

z

Do not separate the glide sheet and top cush-

ion. If they are separated, check that the glide 

sheet is not turned inside out and then insert 

the top cushion. (Ensure the tag is on the outer 

side of the glide sheet.)

Loops (two corners)

String

Glide sheet

Tag

Remove the belts of the 

air cell from the rubber 

loops of the bottom cover.

Remove the belts of the 

air cell from the loops 

of the glide sheet.

Rubber loop

Belt

Belt

Loop

5

 MAINTENANCE AND INSPECTION

Содержание cococia Riraku flow

Страница 1: ...ease contact Paramount Bed or your distributor Intended Use cococia Riraku flow is an air mattress with an internal pump which is designed to disperse the body pressure to prevent pressure ulcers in m...

Страница 2: ...attress and pump and do not allow a person s weight to be applied through their hips or knees The coupling may become disconnected resulting in air leak age and the body pressure not being dispersed e...

Страница 3: ...mp have an expected service life of five years To maintain the condition of the pump have the pump serviced regularly according to the schedule recom mended Do not use unapproved accessories or attemp...

Страница 4: ...nap button Product ID label affixed to side Position sensor cable Pump installation position Sheet stopper Coupling ports Operation panel The product code is printed on the back side Top Bottom Instru...

Страница 5: ...d to the top bottom head end and foot end Failure to do so will result in not being able to disperse the body pressure effectively z Place the product directly on the bed and do not place anything oth...

Страница 6: ...so Orient the plug so that the notch is that the notch is facing downward facing downward and insert it and insert it 3 3 Turn the connector cap Turn the connector cap clockwise clockwise 9 9 Turn the...

Страница 7: ...zippers two locations and pull the operation panel cable and bed link cable out of the cover from the cor ner of the cover between the zippers WARNING z Check that the pump and couplings upper and low...

Страница 8: ...e instruction manual of the bed to learn how to operate the bed and ensure you understand how to safely and properly use the bed before operating it Note 1Make the bottom of the bed flat horizontal 2R...

Страница 9: ...e bed from the outlet and turn the battery off 3Orient the plug without warning label of the bed link cable with its protrusion facing up and insert it into the air mattress outlet of the bed 4Pass th...

Страница 10: ...other than as sheet stoppers 5 Making the bed Towel mattress pad etc 4FUNCTIONS AND OPERATING PROCEDURES 1 Names and functions of operation panel parts 1Max inflate mode button page 13 2Power lamp Li...

Страница 11: ...de is set to ON The pump sound will be loud and the Preparing indication on the operation panel will be lit during fast inflation Functions retained after power off z The function settings immediately...

Страница 12: ...ying position and body shape in addition to the weight during automatic softness control Note z The table for Manual softness control shows the recommended softness level correspond ing to the recomme...

Страница 13: ...SF cells z When back support is disabled all SF cells are disabled if the back angle becomes 45 or more SF sensor off mode If you insert the power plug into an outlet while pressing the Setting button...

Страница 14: ...anual max inflate mode turns on the Max inflate mode button lights in blue and the status display screen displays the remaining time indication How to turn manual max inflate mode off Press the Max in...

Страница 15: ...the type of bed and BED Navi used For details contact Paramount Bed or your distributor 1 button Displays the home screen 2AIR MATTRESS button Displays the AIR MATTRESS home screen see page 15 The fig...

Страница 16: ...al softness control operation 4Preparing indication This is lit during internal pressure adjustment with fast infla tion 5SMALL FLOW SETTING button Changes the small flow setting 6SF cell status displ...

Страница 17: ...F cell adjustment button to enable or disable the SF cell GO BACK RETURN button Confirms the settings and closes the INFLATE SETTING screen SF cell adjustment buttons Touch each SF cell adjustment but...

Страница 18: ...he AIR MATTRESS home screen while automatic soft ness control is set 8MANUAL button Sets the manual softness control function is displayed in the AIR MATTRESS home screen while manual softness control...

Страница 19: ...mold generation z If blood urine sebum and other stains are present use neutral detergent or soap to clean them and do not leave them If such stains are left as is the cover may deteriorate or develop...

Страница 20: ...s attached or no filter is attached insects or dust may enter the inside of the pump possibly causing malfunction This also may cause failure Take out Pump side Filter Filter case Press hook How to cl...

Страница 21: ...uplings secured to the pump properly If they are not secured properly see 3 Connecting the pump and mattress on page 6 and then reconnect them Cable connectors pump Are the cable and AC adapter connec...

Страница 22: ...hile checking that there is no con tact with the mattress base 10 Is the user of the product a patient with a prominent bone projection hump back or lower limb contracture If a user with a prominent b...

Страница 23: ...the length of the mattress Have the user sit down on the side of the bed in a position inside of the sensor detection range 13 Is the back section of the bed raised When the back angle of the bed is 2...

Страница 24: ...it with the CPR zipper side facing the underside of the pump side Pump cover Put into the widest part of the spacer pad Bed link cable Operation panel accessory separately sold Instruction manual Pla...

Страница 25: ...roduct code RC 2030W Protection code IEC 60529 IPx4 Size cm Width x Length x Thickness 5 6 x 5 6 x 17 including hook Product weight kg 0 2 Cable length cm 200 Materials Case ASA resin Hook POM resin S...

Страница 26: ...n China Bed link cable Symbol Description Do not pinch or pull the bed link cable Bed link compatible Refer to the instruction manual Mark for restriction of the use of hazardous substances in electri...

Страница 27: ...max inflate mode Automatic softness control Small flow function Preparing SF cell status display Symbol Description This product is an accessory for medical devices and such accessory is required to b...

Страница 28: ...f the bed Use of RF communication devices within a distance of 30 cm 12 in may cause a decline in the performance of the bed n The bed cannot be used adjacent to other devices or in a stacked state If...

Страница 29: ...lity environment The customer or user of the bed must use this bed in an environment specified here Immunity test IEC 60601 1 2 Test level Compliance level Electrostatic discharge ESD IEC 61000 4 2 8...

Страница 30: ...electromag netic field IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz 80 amplitude modulation 1 kHz 3 V m 80 MHz to 2 7 GHz 80 amplitude modulation 1 kHz a Field strengths from fixed transmitters such as thos...

Страница 31: ...61000 4 3 27 V m 385 MHz 27 V m 380 to 390 TETRA400 Pulse modu lation 18 Hz 1 8 0 3 28 V m 450 MHz 28 V m 430 to 470 GMRS460 FRS460 FM 5 kHz devi ation 1 kHz sine 2 0 3 9 V m 710 MHz 9 V m 704 to 787...

Страница 32: ...performance of the product in stock for at least 6 years after the discontinuation of production 3 Questions regarding after sales service Contact Paramount Bed see below Disposal When disposing of t...

Отзывы: