Using the air mattress

 

z

Take  care  to  prevent  the  drain  tube,  etc.  from  getting 

caught in the gap between the product and the bed, bed 

side-rail, etc.

The drain tube, etc. may come off or break.

 

z

Do not operate this product with part of the body placed 

in a gap between the product and the bed, bed side-rail, 

etc.

The body part may get caught in the gap, possibly causing 

injury.

 

z

For the bed or other products used in combination with 

the  product,  use  suitable  products  specified  by  Para-

mount Bed (see page 24).

Otherwise, unintended gaps may be created, interference 

between products may occur, and stability may be lowered, 

resulting in injury or damage to the product.

 

z

Do not use the product for infants or small children (aged 

12 or under).

There is a risk of injury or suffocation due to pressure caused 

by a part of the body getting caught in the gap between the 

product and a bed side-rail, etc.

 

z

Use a mattress that is the same width and length as the 

bed.

Unintended gaps may be created and part of the body may 

get caught, resulting in injury.

 

z

Do not move the product while a person is on it.

The person may fall off the product resulting in injury or the 

product may be damaged.

 

z

Do not stand or jump on the product.

You may fall or the product may tip over resulting in injury or 

damage to the product.

 

z

Use the product according to the user’s medical condi-

tion.

Depending on the user’s medical condition, the user’s medi-

cal condition could worsen by operating the product.

 

z

Do not use the product for a person lighter or heavier 

than the specified weight range (20 to 180 kg).

Effective body pressure dispersion will not be possible.

 

z

Multiple  persons  should  not  get  on  the  product  at  the 

same time.

The body pressure may not be dispersed effectively, causing 

pressure ulcers or damage to the product.

 

z

Do not lie face down for a long time.

There is a risk of suffocation.

1

 SAFETY PRECAUTIONS

 

z

Do not apply a strong impact to the pump and cable con-

nectors, and do not allow a person’s weight to be applied 

through their hips or knees.

There is a risk of an electric shock or fire due to damage or 

deformation of the pump.

 

z

Do not apply a strong shock to the part connecting the 

mattress and pump, and do not allow a person’s weight 

to be applied through their hips or knees.

The coupling may become disconnected resulting in air leak-

age and the body pressure not being dispersed effectively.

 

z

Do not get the bed link cable caught in the movable parts 

of the bed, run over them with the bed, apply excessive 

force to them, or place a heavy object on them.

They may be damaged, resulting in an electric shock or fire. 

Request repair (or replacement) of damaged cables.

 

z

Install electrical equipment so the bed link cable do not 

get caught in the casters or movable parts of the bed. 

Damage to the cables may cause electric shock or fire.

 

z

Do not drop the pump on the floor, hit it against a wall, 

forcefully pull the cable connected to the pump, etc.

Doing so may cause an electric shock or fire due to damage.

 

z

Do  not  allow  operation  by  a  user,  especially  person 

deemed  incapable  of  understanding  the  operation  (a 

child aged 12 or under or a person with dementia).

Part of the body may get caught in a gap resulting in injury, or 

the body pressure may not be dispersed effectively resulting 

in pressure ulcers.

 

z

If emergency response such as cardiac massage is re-

quired, use the CPR function (see page 17).

Otherwise, immediate action cannot be taken and the symp-

toms may worsen.

 

z

Do not lean the product against a wall, etc.

There is a risk of an electric shock or fire due to the product 

falling.

 

z

Do not place the product in a strong electromagnetic en-

vironment such as an MRI.

• The product may malfunction or fail to work.

• The product needs special precautions regarding EMC and 

needs to be installed according to the EMC information pro-

vided.

 

z

Do not smoke on the product.

Doing so may cause fire.

WARNING

The safety precautions described in this instruction manual are intended to ensure safe use of the 

product and prevent personal injury and property damage. Precautions are classified into “WARN-

ING” and “CAUTION,” which indicate the degree of hazard or injury that could result from improper 

use. They are all important safety precautions that must be strictly observed.

WARNING

Indicates that improper handling by ignoring this indication may cause death or serious injury 

(broken bones, pressure being applied, or paralysis).

CAUTION

Indicates that improper handling by ignoring this indication may cause minor injuries (such as 

bruises, scratches, or cuts) or property damage.

Содержание cococia Riraku flow

Страница 1: ...ease contact Paramount Bed or your distributor Intended Use cococia Riraku flow is an air mattress with an internal pump which is designed to disperse the body pressure to prevent pressure ulcers in m...

Страница 2: ...attress and pump and do not allow a person s weight to be applied through their hips or knees The coupling may become disconnected resulting in air leak age and the body pressure not being dispersed e...

Страница 3: ...mp have an expected service life of five years To maintain the condition of the pump have the pump serviced regularly according to the schedule recom mended Do not use unapproved accessories or attemp...

Страница 4: ...nap button Product ID label affixed to side Position sensor cable Pump installation position Sheet stopper Coupling ports Operation panel The product code is printed on the back side Top Bottom Instru...

Страница 5: ...d to the top bottom head end and foot end Failure to do so will result in not being able to disperse the body pressure effectively z Place the product directly on the bed and do not place anything oth...

Страница 6: ...so Orient the plug so that the notch is that the notch is facing downward facing downward and insert it and insert it 3 3 Turn the connector cap Turn the connector cap clockwise clockwise 9 9 Turn the...

Страница 7: ...zippers two locations and pull the operation panel cable and bed link cable out of the cover from the cor ner of the cover between the zippers WARNING z Check that the pump and couplings upper and low...

Страница 8: ...e instruction manual of the bed to learn how to operate the bed and ensure you understand how to safely and properly use the bed before operating it Note 1Make the bottom of the bed flat horizontal 2R...

Страница 9: ...e bed from the outlet and turn the battery off 3Orient the plug without warning label of the bed link cable with its protrusion facing up and insert it into the air mattress outlet of the bed 4Pass th...

Страница 10: ...other than as sheet stoppers 5 Making the bed Towel mattress pad etc 4FUNCTIONS AND OPERATING PROCEDURES 1 Names and functions of operation panel parts 1Max inflate mode button page 13 2Power lamp Li...

Страница 11: ...de is set to ON The pump sound will be loud and the Preparing indication on the operation panel will be lit during fast inflation Functions retained after power off z The function settings immediately...

Страница 12: ...ying position and body shape in addition to the weight during automatic softness control Note z The table for Manual softness control shows the recommended softness level correspond ing to the recomme...

Страница 13: ...SF cells z When back support is disabled all SF cells are disabled if the back angle becomes 45 or more SF sensor off mode If you insert the power plug into an outlet while pressing the Setting button...

Страница 14: ...anual max inflate mode turns on the Max inflate mode button lights in blue and the status display screen displays the remaining time indication How to turn manual max inflate mode off Press the Max in...

Страница 15: ...the type of bed and BED Navi used For details contact Paramount Bed or your distributor 1 button Displays the home screen 2AIR MATTRESS button Displays the AIR MATTRESS home screen see page 15 The fig...

Страница 16: ...al softness control operation 4Preparing indication This is lit during internal pressure adjustment with fast infla tion 5SMALL FLOW SETTING button Changes the small flow setting 6SF cell status displ...

Страница 17: ...F cell adjustment button to enable or disable the SF cell GO BACK RETURN button Confirms the settings and closes the INFLATE SETTING screen SF cell adjustment buttons Touch each SF cell adjustment but...

Страница 18: ...he AIR MATTRESS home screen while automatic soft ness control is set 8MANUAL button Sets the manual softness control function is displayed in the AIR MATTRESS home screen while manual softness control...

Страница 19: ...mold generation z If blood urine sebum and other stains are present use neutral detergent or soap to clean them and do not leave them If such stains are left as is the cover may deteriorate or develop...

Страница 20: ...s attached or no filter is attached insects or dust may enter the inside of the pump possibly causing malfunction This also may cause failure Take out Pump side Filter Filter case Press hook How to cl...

Страница 21: ...uplings secured to the pump properly If they are not secured properly see 3 Connecting the pump and mattress on page 6 and then reconnect them Cable connectors pump Are the cable and AC adapter connec...

Страница 22: ...hile checking that there is no con tact with the mattress base 10 Is the user of the product a patient with a prominent bone projection hump back or lower limb contracture If a user with a prominent b...

Страница 23: ...the length of the mattress Have the user sit down on the side of the bed in a position inside of the sensor detection range 13 Is the back section of the bed raised When the back angle of the bed is 2...

Страница 24: ...it with the CPR zipper side facing the underside of the pump side Pump cover Put into the widest part of the spacer pad Bed link cable Operation panel accessory separately sold Instruction manual Pla...

Страница 25: ...roduct code RC 2030W Protection code IEC 60529 IPx4 Size cm Width x Length x Thickness 5 6 x 5 6 x 17 including hook Product weight kg 0 2 Cable length cm 200 Materials Case ASA resin Hook POM resin S...

Страница 26: ...n China Bed link cable Symbol Description Do not pinch or pull the bed link cable Bed link compatible Refer to the instruction manual Mark for restriction of the use of hazardous substances in electri...

Страница 27: ...max inflate mode Automatic softness control Small flow function Preparing SF cell status display Symbol Description This product is an accessory for medical devices and such accessory is required to b...

Страница 28: ...f the bed Use of RF communication devices within a distance of 30 cm 12 in may cause a decline in the performance of the bed n The bed cannot be used adjacent to other devices or in a stacked state If...

Страница 29: ...lity environment The customer or user of the bed must use this bed in an environment specified here Immunity test IEC 60601 1 2 Test level Compliance level Electrostatic discharge ESD IEC 61000 4 2 8...

Страница 30: ...electromag netic field IEC 61000 4 3 3 V m 80 MHz to 2 7 GHz 80 amplitude modulation 1 kHz 3 V m 80 MHz to 2 7 GHz 80 amplitude modulation 1 kHz a Field strengths from fixed transmitters such as thos...

Страница 31: ...61000 4 3 27 V m 385 MHz 27 V m 380 to 390 TETRA400 Pulse modu lation 18 Hz 1 8 0 3 28 V m 450 MHz 28 V m 430 to 470 GMRS460 FRS460 FM 5 kHz devi ation 1 kHz sine 2 0 3 9 V m 710 MHz 9 V m 704 to 787...

Страница 32: ...performance of the product in stock for at least 6 years after the discontinuation of production 3 Questions regarding after sales service Contact Paramount Bed see below Disposal When disposing of t...