19.2.1
Change / Enter Reference Value
For certain applications it may be necessary to store a specific reference dimension. This section explains the
procedure for changing the reference dimension to the example value of 1150.0 mm:
Step 1
Step 2
Figure 73: Change/enter calibration unit 1
Figure 74: Change/enter calibration unit 2
Press and hold
ENTER
+
SET
simultaneously
→
“
rES
” and “
0.1
” appear flashing alternately
Press
ENTER
→
“
corFAc
”
and “
1.00000
”
appear
flashing alternately
Step 3
Step 4
Figure 75: Change/enter calibration unit 3
Figure 76: Change/enter calibration unit 4
Press
ENTER
→
The display alternates between
flashing “
reF
” and the originally stored reference
value
→
here in the example “
o01000.0
”.
Press
SET
to select the digit to be changed (the
flashing digit can be modified) and set the
numerical value with the
ABS/INCR
button.
Step 5
Figure 77: Change/enter calibration unit 5
Now press the
ENTER
button 2 times to exit the
programming mode (normal operation).
To check, hold down
ENTER
+
SET
simultaneously
→
The new calibration value “1150.0 mm” appears
in the normal mode (see
Then perform the operation described in section
with the new reference dimension “1150.0
mm” to complete the operation.