background image

- 4 - 

 
 
 

FAILURE OF INDICATORS OR CONTROLS DURING THE 
STARTUP PROCEDURE MAY INDICATE THE NEED FOR 
INSTRUMENT SERVICE. THE INSTRUMENT SHOULD 
NOT BE RETURNED TO SERVICE UNTIL PROPERLY 
OPERATIONAL. 

 
 
 
 

DO NOT USE COMPRESSED AIR FOR BREATHING 
UNLESS THE MONITOR IS DISPLAYING GREEN LIGHTS 
FOR AIR QUALITY INDICATOR AND ALARM STATUS 
INDICATOR. 

 

6.0

 

RESPONSE CHECK 

 
6.1

 

Moisten the sponge in the calibration connector 
as described in the calibration section 

Preparing the Calibration Connector

 

6.2

 

The following items are required for response 
check of the PanBlast

TM

 AirScan Mobile CO 

Monitor: 

 

 

10 ppm carbon monoxide test gas  

 

Part No# BAC-AF-0270-00 

 
 

Calibration valve assembly 

 

Part No# BAC-AF-0269-00 (See Figure 4) 

 
 

DO NOT CHECK THE PANBLAST

TM

 AIRSCAN 

MOBILE CO MONITOR UNTIL IT HAS BEEN IN 
OPERATION ON COMPRESSED AIR FOR AT LEAST 
60 MINUTES TO STABILZE THE SENSOR. 

 
 

Response checking of the PanBlast

TM

 AirScan 

Mobile 

 CO Monitor is accomplished by 

applying 10 PPM CARBON MONOXIDE and 
confirming that the AIR QUALITY indicator turns 
RED. TESTING DOES NOT INVOLVE ANY 
ADJUSTMENT OF THE INSTRUMENT CONTROLS. 

 

6.3

 

The following instructions cover the steps to 
check the PanBlast

TM

 AirScan Mobile CO Monitor: 

 

1)

 

On the calibration valve assembly, turn the 
needle valve clockwise to the closed 
position. 

 
2)

 

Attach the calibration valve assembly to 
the 10 ppm carbon monoxide test gas 
bottle. 

 
3)

 

Move the instrument RUN/CAL switch to 
the CAL position. 

 
4)

 

Attach the calibration connector on the 
calibration valve assembly to the 
instrument test gas connector. Initiate gas 
flow into the instrument by gently turning 
the needle valve of calibration valve 
assembly to the open position. Monitor 
the flow rate to confirm GREEN LED 
illumination on the flow bar graph 
indicator. 

 

 Allow the test gas to flow for 30 seconds, and 
then confirm the internal air quality indicator 
#3, Figure 3 is RED.  If the indicator is GREEN 
the instrument should be taken out of service 
and calibrated. 

 

5)

 

Turn off the feed valve of the calibration 
valve assembly to the closed position and 

remove the test gas from the instrument 
test gas port. 

 
6)

 

Move the RUN/CAL switch to the RUN 
position. 

 

6.4

 

Check and confirm that the flow rate returns to 
the GREEN condition. 

 

6.5

 

Monitor the air quality indicator and confirm it 
has turned back to GREEN, indicating proper 
operation. 

 

7.0

 

OPERATING MODE 

 

7.1

 

Operation that is not continuous will require 
full start-up each time the instrument is used.  
Response tests should be made subsequent to 
the initial startup requirements, page 5 of this 
manual.

 

 

7.2

 

After an initial startup of a PERMANENT 
installation, it is recommended that the 
instrument be TESTED the next day and then 
weekly for the first month. A minimum of 
monthly checks should be established 
thereafter.

 

 

7.3

 

Sensor  life  will  vary.  In  almost  all  cases  the 
sensor will remain effective for one or more 
years. Instruments requiring frequent 
calibration usually indicate the need to 
schedule a sensor replacement. 

 

7.4

 

In normal operation the lid of the instrument 
should be closed except during the calibration 
activity. Care should be exercised during 
calibration to avoid contaminants entering the 
instrument. 

 

7.5

 

The exterior of the instrument may be cleaned 
with a mild soap and water solution. 

 

7.6

 

Temperature excesses should be avoided to 
prolong the life of the sensor. Temperature of 
the sensor cell must be maintained between 4°C 
(40

˚

F) and 48°C (120

˚

F). 

 

7.7

 

The sensor function and overall life will be 
impacted if sample air is extremely dry and 
operations are continuous. 

 

7.8

 

The PanBlast

TM

 AirScan Mobile CO Monitor will 

initiate a low-pressure alarm if the air 
compressor is turned off with the instrument 
still powered. This is typical of operations with 
idle shifts. The low-pressure alarm can be 
silenced with the mode selector switch outlined 
on page 8 of this manual. 

 

8.0

 

CALIBRATION 

 

8.1

 

The following items are required for calibration 
of the PanBlast

TM

 AirScan Mobile CO Monitor, 

they are not supplied with the base unit and are 
optional accessories: - 

 

1

 

CO test gas Part No# BAC-AF-0270-00 

 
2

 

CO free air test gas Part No# BAC-AF-

0271-00 

 
3

 

CO calibration valve Part No# BAC-AF-

0269-00 

 
4

 

Small screwdriver commercially available 

 

DANGER 

DANGER 

Содержание ZVP-PC-0080-00

Страница 1: ...PRODUCT OPERATING MANUAL PANBLASTTM AIRSCAN MOBILE CO MONITOR Manual Number ZVP PC 0070 00...

Страница 2: ...SECTION 1 GENERAL INFORMATION 2 INTRODUCTION 3 INSTALLATION 4 OPERATION 5 INITIAL SET UP 6 RESPONSE CHECK 7 OPERATING MODE 8 CALIBRATION 9 TROUBLESHOOTING GUIDE 10 PARTS LISTING...

Страница 3: ...ectrical trades personnel 1 3 4 Under no circumstances should any safety interlocks lockouts or features be altered or disabled in any way 1 3 5 All equipment must be isolated from the compressed air...

Страница 4: ...t off of fixed installations should be provided Sample line volume is critical keep line as short as possible 3mm 1 8 OD x 1 5mm 1 16 ID Polyurethane tubing is commonly used as the sample line althoug...

Страница 5: ...put port provides the capability to introduce calibration gases to the instrument sensor when the calibration Run valve is in the calibration mode 8 Faceplate door screw 9 Internal sample flow regulat...

Страница 6: ...assembly to the closed position and remove the test gas from the instrument test gas port 6 Move the RUN CAL switch to the RUN position 6 4 Check and confirm that the flow rate returns to the GREEN co...

Страница 7: ...edle knob of the calibration valve assembly to open Monitor the flow bar graph display to ensure GREEN condition and allow the impurity free gas to flow for at least one minute Check the air quality i...

Страница 8: ...board has two 2 12VDC output Molex connectors J3 and J5 shown in Figure 5 One is used to power the internal alarm horn the other is unused and can power an external low current alarm device Maximum cu...

Страница 9: ...sor loosen the screw 8 Figure 3 and open the faceplate When the enclosure is open follow the directions listed below 1 Carefully remove the 2 screws from the sensor block and lift the block off the se...

Страница 10: ...Contact manufacturer for service Alarm horn will not sound Faulty or unplugged horn Plug horn in Replace horn Bad air Air quality indicator is RED Check intake for contamination Out of calibration Re...

Страница 11: ...t Parts Stock Code Description Qty BAC AF 0270 00 CO Test Gas 1 BAC AF 0271 00 Impurity Free Air Test Gas 1 BAC AF 0269 00 Calibration Valve Assembly 1 BAC AF 0341 00 AirScan Sensor 1 BAC AF 0342 00 F...

Отзывы: