- 4 -
FAILURE OF INDICATORS OR CONTROLS DURING THE
STARTUP PROCEDURE MAY INDICATE THE NEED FOR
INSTRUMENT SERVICE. THE INSTRUMENT SHOULD
NOT BE RETURNED TO SERVICE UNTIL PROPERLY
OPERATIONAL.
DO NOT USE COMPRESSED AIR FOR BREATHING
UNLESS THE MONITOR IS DISPLAYING GREEN LIGHTS
FOR AIR QUALITY INDICATOR AND ALARM STATUS
INDICATOR.
6.0
RESPONSE CHECK
6.1
Moisten the sponge in the calibration connector
as described in the calibration section
Preparing the Calibration Connector
.
6.2
The following items are required for response
check of the PanBlast
TM
AirScan Mobile CO
Monitor:
10 ppm carbon monoxide test gas
Part No# BAC-AF-0270-00
Calibration valve assembly
Part No# BAC-AF-0269-00 (See Figure 4)
DO NOT CHECK THE PANBLAST
TM
AIRSCAN
MOBILE CO MONITOR UNTIL IT HAS BEEN IN
OPERATION ON COMPRESSED AIR FOR AT LEAST
60 MINUTES TO STABILZE THE SENSOR.
Response checking of the PanBlast
TM
AirScan
Mobile
CO Monitor is accomplished by
applying 10 PPM CARBON MONOXIDE and
confirming that the AIR QUALITY indicator turns
RED. TESTING DOES NOT INVOLVE ANY
ADJUSTMENT OF THE INSTRUMENT CONTROLS.
6.3
The following instructions cover the steps to
check the PanBlast
TM
AirScan Mobile CO Monitor:
1)
On the calibration valve assembly, turn the
needle valve clockwise to the closed
position.
2)
Attach the calibration valve assembly to
the 10 ppm carbon monoxide test gas
bottle.
3)
Move the instrument RUN/CAL switch to
the CAL position.
4)
Attach the calibration connector on the
calibration valve assembly to the
instrument test gas connector. Initiate gas
flow into the instrument by gently turning
the needle valve of calibration valve
assembly to the open position. Monitor
the flow rate to confirm GREEN LED
illumination on the flow bar graph
indicator.
Allow the test gas to flow for 30 seconds, and
then confirm the internal air quality indicator
#3, Figure 3 is RED. If the indicator is GREEN
the instrument should be taken out of service
and calibrated.
5)
Turn off the feed valve of the calibration
valve assembly to the closed position and
remove the test gas from the instrument
test gas port.
6)
Move the RUN/CAL switch to the RUN
position.
6.4
Check and confirm that the flow rate returns to
the GREEN condition.
6.5
Monitor the air quality indicator and confirm it
has turned back to GREEN, indicating proper
operation.
7.0
OPERATING MODE
7.1
Operation that is not continuous will require
full start-up each time the instrument is used.
Response tests should be made subsequent to
the initial startup requirements, page 5 of this
manual.
7.2
After an initial startup of a PERMANENT
installation, it is recommended that the
instrument be TESTED the next day and then
weekly for the first month. A minimum of
monthly checks should be established
thereafter.
7.3
Sensor life will vary. In almost all cases the
sensor will remain effective for one or more
years. Instruments requiring frequent
calibration usually indicate the need to
schedule a sensor replacement.
7.4
In normal operation the lid of the instrument
should be closed except during the calibration
activity. Care should be exercised during
calibration to avoid contaminants entering the
instrument.
7.5
The exterior of the instrument may be cleaned
with a mild soap and water solution.
7.6
Temperature excesses should be avoided to
prolong the life of the sensor. Temperature of
the sensor cell must be maintained between 4°C
(40
˚
F) and 48°C (120
˚
F).
7.7
The sensor function and overall life will be
impacted if sample air is extremely dry and
operations are continuous.
7.8
The PanBlast
TM
AirScan Mobile CO Monitor will
initiate a low-pressure alarm if the air
compressor is turned off with the instrument
still powered. This is typical of operations with
idle shifts. The low-pressure alarm can be
silenced with the mode selector switch outlined
on page 8 of this manual.
8.0
CALIBRATION
8.1
The following items are required for calibration
of the PanBlast
TM
AirScan Mobile CO Monitor,
they are not supplied with the base unit and are
optional accessories: -
1
)
CO test gas Part No# BAC-AF-0270-00
2
)
CO free air test gas Part No# BAC-AF-
0271-00
3
)
CO calibration valve Part No# BAC-AF-
0269-00
4
)
Small screwdriver commercially available
DANGER
DANGER