- 7 - 

connectors, see Figure 5. There is also a 3 pin 
Molex connector (J1) Figure 5 used to supply 
power from the internal transformer or external 
16 – 18VAC source. 

 
 
 
 

When the circuit board is powered with external DC 
the transformer must be disconnected from (J1) 
transformer connector. 

 

8.10

 

SENSOR REPLACEMENT 

 

Sensor failure will be indicated by increasing 
failure frequency of user response test. The 
following procedure should be observed in 
replacing the sensor. 
 
To access the PanBlast

TM

 AirScan Mobile CO 

Monitor Sensor, loosen the screw #8, Figure 3 
and open the faceplate. When the enclosure is 
open follow the directions listed below: 
 
1

 

Carefully remove the 2 screws from the 
sensor block and lift the block off the 
sensor, Figure 5. 

 
2

 

Remove the senor, which is plugged into 
the circuit board. 

 
3

 

Orient and install the new sensor on 
circuit board.  Align the tab on the sensor 
Figure 7 with the tab indication on the 
circuit board Figure 5, page 8. 

 
4

 

Reinstall the sensor block retaining screws 
taking care not to over tighten.  

  
 
 
 
 
 
 
 
 
 
 
 
 

Figure 7 

 

A new sensor should be operated for several 
hours prior to initial calibration. Calibration 
of a new sensor should be completed as 
outlined on page 6 of this manual. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 

 

 

WARNING 

Sensor Orientation 

Tab 

Содержание ZVP-PC-0080-00

Страница 1: ...PRODUCT OPERATING MANUAL PANBLASTTM AIRSCAN MOBILE CO MONITOR Manual Number ZVP PC 0070 00...

Страница 2: ...SECTION 1 GENERAL INFORMATION 2 INTRODUCTION 3 INSTALLATION 4 OPERATION 5 INITIAL SET UP 6 RESPONSE CHECK 7 OPERATING MODE 8 CALIBRATION 9 TROUBLESHOOTING GUIDE 10 PARTS LISTING...

Страница 3: ...ectrical trades personnel 1 3 4 Under no circumstances should any safety interlocks lockouts or features be altered or disabled in any way 1 3 5 All equipment must be isolated from the compressed air...

Страница 4: ...t off of fixed installations should be provided Sample line volume is critical keep line as short as possible 3mm 1 8 OD x 1 5mm 1 16 ID Polyurethane tubing is commonly used as the sample line althoug...

Страница 5: ...put port provides the capability to introduce calibration gases to the instrument sensor when the calibration Run valve is in the calibration mode 8 Faceplate door screw 9 Internal sample flow regulat...

Страница 6: ...assembly to the closed position and remove the test gas from the instrument test gas port 6 Move the RUN CAL switch to the RUN position 6 4 Check and confirm that the flow rate returns to the GREEN co...

Страница 7: ...edle knob of the calibration valve assembly to open Monitor the flow bar graph display to ensure GREEN condition and allow the impurity free gas to flow for at least one minute Check the air quality i...

Страница 8: ...board has two 2 12VDC output Molex connectors J3 and J5 shown in Figure 5 One is used to power the internal alarm horn the other is unused and can power an external low current alarm device Maximum cu...

Страница 9: ...sor loosen the screw 8 Figure 3 and open the faceplate When the enclosure is open follow the directions listed below 1 Carefully remove the 2 screws from the sensor block and lift the block off the se...

Страница 10: ...Contact manufacturer for service Alarm horn will not sound Faulty or unplugged horn Plug horn in Replace horn Bad air Air quality indicator is RED Check intake for contamination Out of calibration Re...

Страница 11: ...t Parts Stock Code Description Qty BAC AF 0270 00 CO Test Gas 1 BAC AF 0271 00 Impurity Free Air Test Gas 1 BAC AF 0269 00 Calibration Valve Assembly 1 BAC AF 0341 00 AirScan Sensor 1 BAC AF 0342 00 F...

Отзывы: