background image

- 5 - 

8.2

 

The following instructions cover the calibration 
of the PanBlast

TM

 AirScan Mobile CO Monitor. 

 

8.3

 

Preparing the Calibration Connector 

 

 

Test gas is completely dry and free of moisture 
unlike normal air.   

 
 

The sponge in the calibration humidifier 
supplies the necessary moisture to the sample 
gas. The sponge must be moist, but not wet.   

 
 

To moisten the sponge: 

 

(Reference Figure 4) 

 

1

 

Press latch #4 to detach humidifier from 
the valve. 

 

2

 

Carefully add a few drops of distilled 
water in the humidifier #2 to moisten the 
sponge. The sponge must not be wet. If 
necessary shake out any excess. 

 

3

 

Attach humidifier #2 to the valve #5. 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

1)

 

3mm (1/8") OD tubing 

2)

 

Humidifier tube 

3)

 

Cal port connector 

4)

 

Latch 

5)

 

Needle valve 

 

Figure 4  

 

8.4

 

Proceed with calibration as outlined below. 

 

 Complete calibration requires use of impurity 
free test gas to establish the correct zero 
reading prior to further calibration. The 
following instructions outline this procedure.  
When impurity free test gas is not available, the 
initial steps to set zero may be skipped and the 
10 ppm carbon monoxide test gas may be used 
to set the 10 ppm response and alarm. 

 

1

 

On the calibration valve assembly, turn 
the needle valve to the closed position. 

 

2

 

Attach the calibration valve assembly to 
the Impurity free test gas bottle. 

 

3

 

Move the instrument RUN/CAL switch to 
the CAL position. 

 

4

 

Connect the calibration valve assembly 
calibration connector to the instrument 
test gas port. 

 

5

 

Initiate gas flow into the instrument by 
slowly turning the needle knob of the 
calibration valve assembly to open.  

Monitor the flow bar graph display to ensure 
GREEN condition and allow the impurity free 
gas to flow for at least one minute. Check the 
air quality indicator to ensure it is GREEN. 
 

6

 

If it is not GREEN adjust the calibration 
adjustment screw to produce a GREEN 
indication (Shown in Figure 3, page 4). 

 

7

 

Remove the impurity free test gas from 
the instrument and remove the calibration 
valve assembly from the test gas bottle. 

 

8

 

Attach the calibration valve assembly to 
the 10ppm carbon monoxide gas bottle 
and reconnect to the instrument test gas 
port. 

 

Allow 10 ppm carbon monoxide test gas to flow 
for at least 20 seconds and check that the air 
quality indicator has turned RED, if not: 

 

9

 

Adjust the calibration adjustment screw 
until the air quality indicator turns RED. 

 

10

 

Reconnect the impurity free test gas.  
Confirm that the instrument returns to a 
GREEN air quality indication. 

 
 

Properly secure the test gases and equipment. 

 

8.5

 

INTERNAL BOARD SET-UP AND CONNECTORS 

 

The PanBlast

TM

 AirScan Mobile CO Monitor is 

factory configured to operate on 220 VAC and 
to alarm without delay when carbon monoxide 
alarms or improper sample pressure occurs.  
Internal controls and connectors allow user 
modification of the factory settings as outlined 
below. 
 
To access the internal controls and connectors 
the faceplate screw #8, Figure 3 must be 
loosened and the faceplate opened. With the 
faceplate fully opened the user has access to 
sample lines, electrical connectors and the 
mode select switch feature. 

 
 

 
 
Revision of the factory set-up and connections must 
be completed by a technician totally familiar with this 
manual and the instrument application. 

WARNING 

Содержание ZVP-PC-0080-00

Страница 1: ...PRODUCT OPERATING MANUAL PANBLASTTM AIRSCAN MOBILE CO MONITOR Manual Number ZVP PC 0070 00...

Страница 2: ...SECTION 1 GENERAL INFORMATION 2 INTRODUCTION 3 INSTALLATION 4 OPERATION 5 INITIAL SET UP 6 RESPONSE CHECK 7 OPERATING MODE 8 CALIBRATION 9 TROUBLESHOOTING GUIDE 10 PARTS LISTING...

Страница 3: ...ectrical trades personnel 1 3 4 Under no circumstances should any safety interlocks lockouts or features be altered or disabled in any way 1 3 5 All equipment must be isolated from the compressed air...

Страница 4: ...t off of fixed installations should be provided Sample line volume is critical keep line as short as possible 3mm 1 8 OD x 1 5mm 1 16 ID Polyurethane tubing is commonly used as the sample line althoug...

Страница 5: ...put port provides the capability to introduce calibration gases to the instrument sensor when the calibration Run valve is in the calibration mode 8 Faceplate door screw 9 Internal sample flow regulat...

Страница 6: ...assembly to the closed position and remove the test gas from the instrument test gas port 6 Move the RUN CAL switch to the RUN position 6 4 Check and confirm that the flow rate returns to the GREEN co...

Страница 7: ...edle knob of the calibration valve assembly to open Monitor the flow bar graph display to ensure GREEN condition and allow the impurity free gas to flow for at least one minute Check the air quality i...

Страница 8: ...board has two 2 12VDC output Molex connectors J3 and J5 shown in Figure 5 One is used to power the internal alarm horn the other is unused and can power an external low current alarm device Maximum cu...

Страница 9: ...sor loosen the screw 8 Figure 3 and open the faceplate When the enclosure is open follow the directions listed below 1 Carefully remove the 2 screws from the sensor block and lift the block off the se...

Страница 10: ...Contact manufacturer for service Alarm horn will not sound Faulty or unplugged horn Plug horn in Replace horn Bad air Air quality indicator is RED Check intake for contamination Out of calibration Re...

Страница 11: ...t Parts Stock Code Description Qty BAC AF 0270 00 CO Test Gas 1 BAC AF 0271 00 Impurity Free Air Test Gas 1 BAC AF 0269 00 Calibration Valve Assembly 1 BAC AF 0341 00 AirScan Sensor 1 BAC AF 0342 00 F...

Отзывы: