283
3.
Use “dual pointer” of the guide laser function to confirm the setting position of the power meter
detector. Set the detector so that the cross indication of the dual pointer radiates on the center of
the detector. (Do not set it to the position where the dot indication and cross indication overlap.)
4.
Press “ADJUST.” in the operation screen.
5.
Press “LSR. CHK.”.
Press “START” in the laser radiation confirmation dialog
to irradiate the laser.
Reference
• The power setting in the selected file is applied to the laser check radiation.
WARNING
•
Be sure to wear protective goggles.
•
During the radiation, the laser energy is concentrated to one point. Use due
caution with long period radiation, it may cause a fire or damage to the object.
6.
Press “Stop” to stop laser radiation.
Without “stop”, the laser radiation automatically stops after about
one minute.
ME-LP400V-OP-6