Chapter 3.  

32

TransPort™ PT878GC User’s Manual

3.

With one hand, mount one transducer into the PI fixture. With the other hand, tighten the transducer hold down 

screw, gradually pushing the transducer down to the pipe surface. Use a wrench to tighten the backing nut to 

prevent loosening due to vibration and thermal expansion. 

IMPORTANT:

Do not use pliers or a wrench on the thumbscrew.

4.

Repeat step 3 for the other transducer.

Содержание PT878GC

Страница 1: ...panametrics com Flow 910 229 B3 August 2021 TransPort PT878GC Portable Gas Flowmeter Instruction Manual...

Страница 2: ......

Страница 3: ...ntains one or more registered trademarks of Baker Hughes Company and its subsidiaries in one or more countries All third party product and company names are trademarks of their respective holders Tran...

Страница 4: ...ii no content intended for this page...

Страница 5: ...Fixture and Transducers 28 3 1 4 1 Surveying the Pipe 28 3 1 4 2 Installing the First Bracket with a Chain or Strap 30 3 1 4 3 Installing the Second Bracket with a Chain or Strap 30 3 1 4 4 Installin...

Страница 6: ...1 14 Listing Files in Chronological Order 87 6 1 Displaying and Configuring Data 89 6 1 1 The Format Option 89 6 1 2 The View Option 91 6 1 3 The Limits Option 92 6 1 4 The Measurement Option 93 6 1 5...

Страница 7: ...137 9 1 Servicing the PT878GC 139 9 1 1 Entering the Service Menu 140 9 1 2 Printing Reports 140 9 1 3 Setting up the Thickness Gauge 143 9 1 4 Measuring Pipe Wall Thickness 144 9 1 4 1 Entering the M...

Страница 8: ...gon 182 11 1 1 10 Pressure Requirements Natural Gas 182 11 1 1 11 Pressure Requirements Steam 182 11 1 1 12 Pressure Requirements Other Gases 182 11 1 1 13 Measurement Parameters 182 11 1 2 Electronic...

Страница 9: ...1 5 Additional Options 185 11 1 5 1 PC Communication 185 11 1 5 2 RS232 to Infrared 185 11 1 5 3 Printer 185 A 1 Menu Maps 187 B 1 Establishing IR Communications with the PT878GC 197 B 1 0 1 Windows...

Страница 10: ...Contents viii TransPort PT878GC User s Manual...

Страница 11: ...nd accessories Figure 1 PT878GC in Use and in Case The TransPort has the ability to store site data in files which can be accessed at a later time Within the Main Menu a set of forms windows asks you...

Страница 12: ...tted transducers onto the pipe or inserted as a substitute pipe section spoolpiece The flowcell must provide mechanical support for the transducers and assure stable conditions for accurate flow measu...

Страница 13: ...es the received signals to a level suitable for the data acquisition circuitry 4 DATA ACQUISITION The data acquisition circuitry digitizes the received signal and stores it in a buffer for processing...

Страница 14: ...ators and receivers When mounted on a pipe they are in acoustic communication with each other so that each transducer can receive ultrasonic signals transmitted by the other transducer Each transducer...

Страница 15: ...pecially designed case Figure 2 The PT878GC and Accessories 2 1 1 Making Electrical Connections Before making measurements with the TransPort you must make all the necessary connections to the unit Th...

Страница 16: ...Chapter 2 6 TransPort PT878GC User s Manual Figure 3 Connection Locations...

Страница 17: ...3 on the previous page Transducer installation is discussed separately in Chapter 3 2 1 1 3 Analog Input Output Connections The TransPort provides one 0 4 20 mA current output and two 4 to 20 mA analo...

Страница 18: ...off If the PT878GC is on the Battery icon in the upper right corner of the screen indicates battery status as shown in Table 2 below IMPORTANT For CE compliance the PT878GC is classified as a battery...

Страница 19: ...ad Immediately upon power up the PT878GC emits a short beep and displays a PCI Loader message It then validates the instrument programming and then displays the Panametrics Sensing logo and the softwa...

Страница 20: ...ower supplies shut down but the PT878GC remains in a standby mode Users can resume taking measurements immediately by pressing the power button Press F3 to cancel the command and return the PT878GC to...

Страница 21: ...ays the measured parameters numeric measurements and both bar and line graphs When you enter a selection on the Main Menu discussed in Chapter 4 Programming Site Data this area displays menu prompts A...

Страница 22: ...functions depending on the task you are performing Stopwatch Calibration Gate Operation Watch is stopped when the gate is closed or runs when it is open See page 3 46 Snapshot To file Indicates that t...

Страница 23: ...ric data 4 arrow keys enable you to move through the menu options Help key enables you to access on line help MENU Menu key enables you to access the Menu Bar ENTER enables you to enter a particular m...

Страница 24: ...keys to navigate to the desired menu and press ENTER Repeat this procedure to access the desired topic within the menu When you have finished using the Help menu Press F1 Back to move back one level P...

Страница 25: ...ries of fixtures The V1 clamping fixture for pipes with diameters between 0 75 and 1 25 in 20 to 30 mm The V4 clamping fixture for pipes with diameters between 1 25 and 4 in 30 to 100 mm The V8 clampi...

Страница 26: ...measurement Specifications assume a fully developed flow profile typically requiring 20 diameters upstream and 10 diameters downstream of straight pipe run In addition pipes 2 in and under require an...

Страница 27: ...um pipe sizes schedules and pressures for use with the PT878GC For air nitrogen oxygen or argon refer to Table 4 For natural gas refer to Table 5 For steam refer to Table 6 1 Find the pipe size of you...

Страница 28: ...5 1 120 36 6 69 21 0 46 14 0 0 22 5 6 0 5 4 100 0 2 1 60 5 1 120 36 6 53 16 2 35 10 7 0 24 6 1 0 5 60 5 1 0 34 8 6 0 5 180 13 4 0 68 17 3 0 5 300 21 7 6 150 0 2 1 60 5 1 90 27 4 72 22 0 54 16 5 0 28...

Страница 29: ...1 7 55 16 8 44 13 4 33 10 1 0 2 1000 68 82 25 66 20 1 49 14 9 14 350 0 38 9 7 0 2 90 7 2 87 26 5 70 21 3 52 15 9 0 50 12 7 0 2 270 19 6 16 400 0 38 9 7 0 2 90 7 2 76 23 2 61 18 9 45 13 7 0 50 12 7 0 2...

Страница 30: ...28 6 100 30 5 80 24 4 60 18 3 0 2 500 34 150 45 7 120 36 6 90 27 4 0 88 22 4 0 5 800 56 2 100 30 5 80 24 4 60 18 3 0 2 1000 68 150 45 7 120 36 6 90 27 4 10 250 0 37 9 4 0 5 200 14 8 85 25 9 68 20 7 5...

Страница 31: ...Velocity ft s m s Single Traverse 3 80 0 22 5 6 0 5 110 8 6 120 36 6 0 3 7 6 0 5 200 14 8 4 100 0 24 6 1 0 5 110 8 6 120 36 6 0 34 8 6 0 5 200 14 8 6 150 0 28 7 2 0 5 110 8 6 120 36 6 0 44 11 2 0 5 20...

Страница 32: ...er side of the pipe for easy transducer installation 150 mm 6 in if you are not using a junction box or 225 mm 9 in if you are using a junction box In general transducers are installed 180 apart on op...

Страница 33: ...tire layout tape mark two circumferential lines along the edges of the wrap as shown in Figure 14 below Figure 14 Measuring and Marking Circumference Lines 4 Now measure the OD and the wall thickness...

Страница 34: ...location as shown in the equations below where ODX is the outside diameter at a given point and WX is the wall thickness at a given point as shown in Figure 3 6 on the previous page 3 Program the valu...

Страница 35: ...sition the half of the clamping fixture with the threaded rods around the pipe as shown in Figure 16 below Orient the fixture in the 3 o clock position on a horizontal pipe 2 Position the mating half...

Страница 36: ...he same origin as shown in Figure 17 below Figure 17 Fixture with Scale Origins Properly Aligned 3 Install the four nuts onto the threaded rods with the convex side of the nut facing the fixture Hand...

Страница 37: ...which in turn applies pressure to the transducer Use a handtight grip to set the transducer in contact with the pipe as shown in Figure 20 below Use a wrench to tighten the backing nut to prevent loo...

Страница 38: ...to an accuracy of 2 mm 1 16 in IMPORTANT Do not use a calculated value or a nominal value for the circumference 2 Tightly wrap the layout tape once around the entire pipe and line up the edges Using t...

Страница 39: ...the zero point as shown in Figure 24 below Place marks on both sides of the circumferential lines made with the layout tape and connect the marks Figure 24 Marking the 9 o Clock Position Make sure to...

Страница 40: ...l and axial location Center the indicator line on the block to line up with the scribed mark see Figure 27 below Figure 27 Positioning the CFG PI Holder Bracket 5 Loosen the transducer hold down screw...

Страница 41: ...t shown in Figure 30 below Figure 30 The Installed CFG PI Fixture with Calculated Spacing 3 1 4 4 Installing the Transducers 1 Check to be sure the second CFG PI holder bracket is correctly positioned...

Страница 42: ...e other hand tighten the transducer hold down screw gradually pushing the transducer down to the pipe surface Use a wrench to tighten the backing nut to prevent loosening due to vibration and thermal...

Страница 43: ...installation The DMP 3 is a clay like compound for all temperature applications If the temperature is over 150 F the PDJ pipe dampening jacket available from Panametrics Sensing with preapplied DMP 3...

Страница 44: ...ere properly the material must also be at a temperature above 50 F 10 C 5 Before more atmospheric condensation can occur on the outside of the pipe roll the DMP 1 dampening material once around the pi...

Страница 45: ...over 10 Reinstall the fixture and transducers on the pipe 11 Lay the second strip of dampening material flat Then cut the strip into two pieces each 4 5 in 12 cm wide 12 Wrap each of these strips arou...

Страница 46: ...stance Between Transducer Faces 3 Lay the first strip of dampening material flat Cut a strip to a width equal to the distance between transducer faces the distance illustrated in Figure 36 above 4 Wra...

Страница 47: ...ile wearing appropriate gloves degrease the surface as shown in Figure 38b above 5 Place a piece of the DMP 3 material on top of the pipe Figure 39a and use the palm of the hand to press it onto the p...

Страница 48: ...g jacket with preapplied DMP 3 As the dampening material dries out over several hours after installation its effectiveness increases The jacket is available in standard pipe sizes from 4 to 12 in 100...

Страница 49: ...couplant Spread it in a thin layer about 6 mm 0 25 in wide on each transducer face 6 Install the transducers into the yokes and tighten the hold down screws until the couplant touches the pipe 7 If y...

Страница 50: ...Chapter 3 40 TransPort PT878GC User s Manual...

Страница 51: ...n a PC and then upload the sites as they are needed This chapter covers entering The Program Menu Transducer pipe and fluid parameters Input output and other setup parameters User functions 4 1 1 Ente...

Страница 52: ...gas transducers and fixture configurations Figure 45 The Transducer Option Window 1 The first prompt asks you to select whether you are using a wetted or a clamp on transducer a Use the and keys to sc...

Страница 53: ...ENTER 4 The next prompt asks for Tw the time delay This parameter is actually the time the transducer signal spends travelling through the transducer and cable The PT878GC calculates the flow rate by...

Страница 54: ...d press ENTER The screen appears similar to Figure 46 below To step through each parameter press the key Refer to Figure A 4 on page A 4 of Appendix A Menu Maps Note Refer to the brochure Sound Speeds...

Страница 55: ...CPVC you have the option of entering the pipe dimensions by a standardized schedule Once you enter the nominal pipe size and schedule identification the PT878GC determines the OD and wall thickness f...

Страница 56: ...n page 7 3 the schedules appear as ANSI b Press the key twice to move the prompt to the Nominal pipe size drop down menu Press ENTER to open the menu Scroll to the desired pipe size and press ENTER to...

Страница 57: ...to confirm your choice Note If your pipe lining is not on the drop down list consult Panametrics Sensing for further information 2 The program follows one of two paths a If you have selected a preprog...

Страница 58: ...s Steam b Scroll to the appropriate entry If you do not see your fluid on the list select Other c Press ENTER to confirm your selection If you have selected a preprogrammed gas the PT878GC supplies th...

Страница 59: ...rrow key to reach the Path window and press ENTER The screen appears similar to Figure 49 below To step through each parameter press the key Refer to Figure A 3 on page A 3 of Appendix A Menu Maps Fig...

Страница 60: ...u enter an invalid entry the PT878GC rejects the entry and displays an error message Note Panametrics Sensing does not recommend using a spacing other than the one calculated by the PT878GC After you...

Страница 61: ...he entry 3 The next prompt asks for the atmospheric pressure a Press ENTER to enter the text box b Use the numeric keys to enter the local atmospheric pressure c Press ENTER to confirm the entry 4 The...

Страница 62: ...nputs window The screen appears similar to Figure 51 below To step through each parameter press the key Figure 51 The Inputs Tab in the Standard Volume Mass Flow Option 1 The first prompt asks if the...

Страница 63: ...lection 5 The next prompt appears if you have selected a fixed temperature The PT878GC asks for the fixed temperature value a Press ENTER to open the text box b Use the numeric keys to enter the known...

Страница 64: ...ree options are available static density saturated steam and steam a Press ENTER to open the drop down menu b Scroll to the appropriate method for your application c Press ENTER to confirm the entry I...

Страница 65: ...and Delta T The default value is 80 and values from 1 to 100 are acceptable a Press ENTER to enter the text box b Use the numeric keys to enter the recommended value c Press ENTER to confirm the entr...

Страница 66: ...ult value is 3 and values from 0 to 10 are acceptable 6 The next prompt enables you to reset the size of the transmit window Normally the PT878GC calculates the size of both transmit and receive windo...

Страница 67: ...ic keys to enter the recommended value c Press ENTER to confirm the entry 2 Next enter the Noise Level Offset of Peak from 0 to 95 and press ENT The default value is 20 a Press ENTER to enter the text...

Страница 68: ...ic keys to enter the recommended value c Press ENTER to confirm the entry 4 If you wish to select or deselect Measure Times only instead of both Skan and Measure times press ENTER at the Measure Only...

Страница 69: ...ntify the peak of the received signal the PT878GC uses the Threshold method in which the peak is identified as the point where the signal crosses a threshold that is a percentage of the maximum signal...

Страница 70: ...er the desired value c Press ENTER to confirm the entry 4 To select or deselect Measure Times only instead of both Skan and Measure times press ENTER at the Measure Only prompt Pressing the key return...

Страница 71: ...t prompt asks you to select whether the desired function is off or general purpose a Press ENTER to open the drop down menu b Scroll to the desired response c Press ENTER to confirm your selection Not...

Страница 72: ...Cancel to leave the window without adding the label 3 The next prompt asks for the Units Symbol a Press ENTER to reopen the text creation window b Repeat the procedure used for the Label on the previo...

Страница 73: ...y to step through these parameters 2 The first prompt asks for the zero input value a Press ENTER to enter the text box b Use the numeric keys to enter the desired value c Press ENTER to confirm the e...

Страница 74: ...to send a current signal to a recording device a Press ENTER to open the drop down menu b Scroll to the desired output from three choices Off 0 20 mA and 4 20 mA c Press ENTER to confirm your selectio...

Страница 75: ...enter the span full scale value for the analog output This value represents the 20 mA output in flow units a Press ENTER to open the text box b Use the numeric keys to enter the desired value c Press...

Страница 76: ...lizer Frequency Test Points Gate Input a Press ENTER to open the drop down menu b Scroll to the desired function c Press ENTER to confirm your selection Note If you select Off you will not be able to...

Страница 77: ...n For Polarity a Press ENTER to open the drop down menu b Scroll to the desired polarity either Low High or High Low c Press ENTER to confirm your selection If you selected Frequency The prompt asks f...

Страница 78: ...o perform mathematical equations on each measurement You can also use any parameter in the meter to calculate a different parameter To enter the User Functions option scroll to the User Functions entr...

Страница 79: ...t the procedure used for the Label on the previous page to create the Units Symbol c When you have completed the entry press F3 OK to confirm the symbol or F2 Cancel to leave the window without adding...

Страница 80: ...OK or a message such as Syntax Error Press F1 to delete the entire function and start again or Press F2 to cancel the function and return to Operate Mode Press F3 Done to confirm the function and retu...

Страница 81: ...lection 3 The following steps depend on whether you select a single value or a table for Kinematic Viscosity If you entered Single The prompt asks for a kinematic viscosity value a Press ENTER to open...

Страница 82: ...ss F3 OK to confirm the table and return to the Correction Factors window You have completed entering data for a Reynolds Correction table To confirm the entries press F3 OK The meter returns to Opera...

Страница 83: ...enabled the Reynolds Correction factor the correction factor should be set to 1 00 Otherwise the typical factor is between 0 5 and 2 00 c Press ENTER to confirm your selection You have completed ente...

Страница 84: ...Repeat steps b and c for the remainder of the table e When you have completed entering values press F3 OK to confirm the table and return to the Correction Factors window You have completed entering...

Страница 85: ...u press the MENU key at the lower right of the PT878GC keypad The Menu Bar replaces the Status Bar at the top of the screen Then press ENTER The screen now appears similar to Figure 71 below While fol...

Страница 86: ...n use the and arrow keys to scroll to a particular site and display information pertaining to that site To open the File submenu within the Site Manager press MENU and then ENTER The screen now appear...

Страница 87: ...below Figure 74 Name Entry Window for a New Site 1 Use the four arrow keys to scroll to the desired letter or number and press ENTER 2 Repeat this procedure until you have created the desired site na...

Страница 88: ...firmation Window The meter returns to Operate Mode with the new site name displayed in the upper left corner of the screen A B C D E F G H I J K L M N O P Q R S T U V W X Y Z 0 1 2 3 4 5 6 7 8 9 New S...

Страница 89: ...is the one you are now closing by opening the second site the PT878GC asks if you want to save the changes to the first site Press F2 No to cancel the changes or Press F3 Yes to confirm and save the c...

Страница 90: ...77 The Save Current Site Window Press F2 No to cancel saving the site or Press F3 Yes to save the site The PT878GC remains in the current window Operate Mode or Site Manager with the current site save...

Страница 91: ...te window then appears Press F2 No to cancel saving the site or Press F3 Yes to save the site The PT878GC remains in Site Manager with the current site saved under both the old and new names 5 1 1 5 R...

Страница 92: ...eting a Site To delete a site in the Site Manager first be sure you have highlighted that site in the left window of the Site Manager Then press MENU scroll to the Delete option and press ENTER The sc...

Страница 93: ...nu to the Site Menu Press ENTER The screen appears similar to Figure 81 below Figure 81 The Site Menu in the Site Manager 2 Press ENTER to select Message which opens the text creation window shown in...

Страница 94: ...capital letters a set of lower case letters and a set of symbols Use all three screens to create a desired message 4 Repeat this procedure for each letter or symbol you wish to add to the message If y...

Страница 95: ...C is searching for an infrared device If the scanner finds no device a window appears with this message and asks if you want to try again If it finds a device it sends another message indicating that...

Страница 96: ...To download a site or meter file from a PC to a PT878GC 1 Check that the Communications option see page 7 15 has been set to the IrDA protocol Note For Windows NT 4 0 check that the QuickBeam softwar...

Страница 97: ...rder Figure 85 The Sort Submenu 5 1 1 14 Listing Files in Chronological Order If you prefer to list your files chronologically by time of creation within the Site Manager press MENU to open the File M...

Страница 98: ...Chapter 5 88 TransPort PT878GC User s Manual...

Страница 99: ...ments Once you have reached a particular window pressing the ENTER button opens the menu for that window as shown in Figure 86 below Figure 86 Display Screen After Pressing SEL 6 1 1 The Format Option...

Страница 100: ...s ENTER to open the drop down list of format choices b Scroll to the appropriate selection c Press ENTER to confirm your choice 2 The second step asks you to choose the number of decimal places to be...

Страница 101: ...on and then press ENTER 2 A drop down menu shows the three formats Scroll to the appropriate selection 3 Press ENTER to confirm your selection The screen immediately displays any changes For example i...

Страница 102: ...to confirm the value 5 Repeat the procedure in step 5 to enter the maximum value A line graph involves two more prompts 6 The next prompt asks for the display interval a Press ENTER to open the text b...

Страница 103: ...measurement data sources while the right column displays English and metric measurement units or with the Diagnostics source shown a list of diagnostics parameters 4 Use the and arrow keys to reach th...

Страница 104: ...s Bar at the top of the screen The Site Menu will be highlighted in the upper left corner Press ENTER or the arrow key The screen now appears similar to Figure 91 below While following the programming...

Страница 105: ...splays the designated number of windows For example Figure 92 below shows the screen displayed in Figure 6 3 on page 6 4 reconfigured for a single view Figure 92 Display Screen Configured for Single V...

Страница 106: ...to move to the FKeys entry on the menu Press ENTER The screen now appears similar to Figure 93 below Figure 93 The FKeys Menu 3 Use the or arrow keys to move to the desired FKey 1 2 or 3 entry on the...

Страница 107: ...s to confirm the assignment and close the window Press F2 No to close the window without changing the key Figure 95 The Display Screen with Modified Softkey To clear a softkey 1 Press MENU 2 From the...

Страница 108: ...n the meter To open the Drive Manager 1 Press MENU 2 From the Site Menu shown in Figure 6 6 on page 6 8 use the or arrow keys to move to the Drive Manager entry on the menu Press ENTER The screen now...

Страница 109: ...s highlighted in the window on the left Then press MENU and scroll to the appropriate option as shown in Figure 97 above Press ENTER The PT878GC then performs the desired action with the file To refre...

Страница 110: ...sends another message indicating that it is uploading the site The meter also displays a message if the upload fails When the upload is complete the meter returns to the Drive Manager The PC holds the...

Страница 111: ...et button in the QuickBeam window as shown in Figure 99 below Figure 99 The QuickBeam Window Open the Send menu in the QuickBeam window enter the Files or Files on Clipboard option scroll to the desir...

Страница 112: ...e Delete option and press ENTER The screen appears similar to Figure 100 below Figure 100 The Delete Confirmation Window Press F2 No to cancel the deletion and return to the File Manager or Press F3 Y...

Страница 113: ...enu 6 1 6 6 Listing Files in Chronological Order If you prefer to list your files chronologically by time of creation within the File Manager press MENU to open the File Menu Then press the arrow key...

Страница 114: ...scroll to the About option on the Site Menu and press ENTER The screen appears similar to Figure 102 below Figure 102 The About Window To find out details about your specific software version press Ne...

Страница 115: ...global settings for the meter that suit their individual preferences The global settings include English or Metric measurement units Battery power Date and time parameters and appearance Screen contr...

Страница 116: ...ameters The key to scroll back to a previously entered parameter The F2 key Cancel or the ESC key to exit an option at any time and return to Operate Mode without changing the parameters 7 1 2 Selecti...

Страница 117: ...eys to reach the prompt a Press ENTER to open the drop down menu b Use the and arrow keys to scroll to the desired choice PSIA KPA Bars MMGH c Press ENTER to confirm the entry To confirm the entries a...

Страница 118: ...from a 1 8 Ahr NiCd battery part number 200 058 or a 3 0 Ahr NiMH battery part number 200 081 Scroll to the battery type you have installed and press ENTER Note If you do not set the correct battery t...

Страница 119: ...ice changes the number to 07 You can scroll from 01 to 12 for the month and from 01 to 31 for the day depending on the number you have selected for the month In either case press ENTER to confirm the...

Страница 120: ...symbols for the date time and decimal You can choose from a dash comma and slash for the date from a period and colon for the time and from a period or comma for the decimal place For each symbol typ...

Страница 121: ...arrow keys to scroll to the 12 hour or 24 hour entry c Press ENTER to confirm your entry A line at the bottom the Date Time Decimal Formats displays how the format and separator selections will appea...

Страница 122: ...p through each parameter press the key Figure 107 Display Tab in the Display Options Window 2 Scroll to the Darker or Lighter box as desired 3 Press the ENTER button repeatedly until the screen has th...

Страница 123: ...splay Options window opens on the Display tab as shown in Figure 7 5 on page 7 12 2 Press the key three times to reach the Backlight Off text box 3 Then press ENTER to enter the box 4 Use the numeric...

Страница 124: ...iation and the IR232 virtual interface You must use IrDA when using an IR printer or transfer application such as QuickBeam Use the and keys to scroll to the desired selection and press ENTER Note For...

Страница 125: ...stop bits The default number is one Use the and keys to scroll to the desired number and press ENTER 7 The final prompt asks you to select either seven or eight data bits The default number is 8 Use...

Страница 126: ...ss ENTER The window now appears similar to Figure 109 below Figure 109 The Resetting Totals Menu 2 Use the or arrow keys to choose the total to be reset 3 Press ENTER to reset the total s The meter re...

Страница 127: ...w now appears similar to Figure 110 below Figure 110 The User Tables Window 2 The first prompt asks you to select the table to be programmed a Press ENTER to open the drop down menu b Use the or arrow...

Страница 128: ...Table ID window can display up to 15 characters If you wish to delete a letter press F1 Delete to erase each letter or symbol from right to left on the label d When you have completed the label press...

Страница 129: ...h then appears in the appropriate slot in the table e Repeat steps b c and d on page 7 21 until you have completed entering data for the table f When you have finished press F3 OK to confirm the entri...

Страница 130: ...screen press the decimal button in the lower left corner of the keypad twice The meter beeps indicating that it has made a screen capture The To Printer option enables you to send a picture of the cu...

Страница 131: ...of memory is assigned to data logging The frequency of the time interval the length of the log run and the number of logs affect the amount of memory required for a particular log For example a log t...

Страница 132: ...y to step through the available parameters The key to scroll back to a previously entered parameter The F2 key Cancel or the ESC key to exit an option at any time and return to Operate Mode without ch...

Страница 133: ...e cursor highlights the File Menu in the upper left corner Use the and keys to scroll to the desired menu and press ENTER to open the menu Log Manager Info AAA LOG 03 01 04 13 23 21 1216 bytes 109568...

Страница 134: ...from the Log Manager press the MENU key and then ENTER The screen appears similar to Figure 115 below Figure 115 File Menu in the Log Manager Log Manager Info AAA LOG 03 01 04 13 23 21 1216 bytes 1095...

Страница 135: ...TER In either case the Create New Log screen appears similar to Figure 116 below Use the four arrow keys to scroll to the desired letter or number and press ENTER Repeat this procedure until you have...

Страница 136: ...e key to scroll to the time box b Two alternatives are available to change a highlighted number Use the numeric keys to enter the desired number Use the or arrow keys to scroll in 1 digit increments t...

Страница 137: ...he first entry The Select Measurement window opens as shown in Figure 119 Figure 119 The Select Measurement Window 2 Scroll to the desired output type 3 Press SEL to confirm your selection General Ns...

Страница 138: ...se specified the cloned log always begins at the current date and time and ends after the period of time used by the original log 5 If you wish scroll to the Measurements window shown in Figure 118 an...

Страница 139: ...8GC shows a message indicating that its infrared sensor is looking for a printer If the sensor finds no printer a window appears with this message If it finds two printers within its range it asks you...

Страница 140: ...Menu in the Log Manager 8 1 4 1 Stopping Pausing a Log To stop a log that is currently pending or running 1 First select the log you wish to pause in the Log Manager shown in Figure 114 2 Then open t...

Страница 141: ...ed logs open the Log menu scroll to the Start All Logs option and press ENTER The PT878GC returns to the Log Manager which displays the logs with a status of Pending or Running 8 1 4 6 Ending All Logs...

Страница 142: ...ew menu from the Log Manager press MENU Scroll to the View menu and press ENTER The screen appears similar to Figure 122 below Figure 122 The View Menu Log Manager Info AAA LOG 03 01 04 13 23 21 1216...

Страница 143: ...the View menu and press ENTER Scroll to the Details option and press ENTER The screen now appears similar to Figure 123 below Figure 123 The Log Details Display Window with List of Measurements in the...

Страница 144: ...lar to Figure 124 below Figure 124 The Log Graph Display Window Note If a log contains more than one measurement you must select which measurement to plot from the list displayed If you wish to alter...

Страница 145: ...raph If you wish to alter the time scale press F2 Time The Enter Time window appears similar to Figure 8 14 below Figure 126 The Enter Time Window 1 Use the arrow keys to scroll to the desired text bo...

Страница 146: ...og Spreadsheet Window Use the and keys to scroll to other columns or the or keys to scroll backward or forward in time To alter the times displayed press F1 Time The Enter Time window shown on the pre...

Страница 147: ...Log Manager screen refreshes with the sites listed in alphabetical order 8 1 6 2 Listing Logs in Chronological Order If you prefer to list your logs chronologically by time of creation within the Log...

Страница 148: ...Chapter 8 138 TransPort PT878GC User s Manual...

Страница 149: ...a variety of functions that they might occasionally require print out reports calibrate the PT878GC run diagnostics set up signal parameters and peak detection define error limits test the meter scre...

Страница 150: ...eters in an option press The key to step through the available parameters The key to scroll back to a previously entered parameter The F2 key Cancel or the ESC key to exit an option at any time and re...

Страница 151: ...keys to scroll to the desired reports and press ENTER Then press F2 Print to print out the desired reports or press F3 Exit to leave the option and return to Operate Mode Figure 131 on the next page s...

Страница 152: ...Chapter 9 142 TransPort PT878GC User s Manual Figure 131 Printout of a Typical Drive Report...

Страница 153: ...le accuracy you can also calibrate the velocity of sound in the pipe material Using the thickness gauge involves three steps 1 Entering the material and sound speed in the Material window on page 144...

Страница 154: ...ow keys to scroll to the desired material c Press ENTER to confirm your selection If you have selected a preprogrammed material you have completed entering data in this window Pressing the key returns...

Страница 155: ...play option shows the thickness measurement in numeric format displayed in Figure 134 below Note Refer to Appendix C Ultrasonic Thickness Gauge Theory of Operation page C 3 to learn how to position an...

Страница 156: ...med values Also some materials such as Teflon or fiberglass do not support acoustic signals To open the Graph window and display the receive signal graphically press the or arrow key until you reach t...

Страница 157: ...Out to zoom out fully Press F3 to toggle between the left and right cursors Press and to move the selected cursor left and right Press to select the graph Press to return to tab navigation Press ESC t...

Страница 158: ...ER to open the text box b Use the numeric keys to enter the length c Press ENTER to confirm the entry 4 Press ENTER Calibrate button to prepare for calibration Then apply couplant B to the block surfa...

Страница 159: ...on To open the Velocity window press the or arrow key until you reach the Velocity tab and press ENTER The screen appears similar to Figure 138 below Figure 138 The Sound Speed Window 1 Scroll to the...

Страница 160: ...value which becomes the new Calculated value Note If you commit the calculated value the settings in the Material window also change from the previous material to Other with the new sound speed To le...

Страница 161: ...open the text box b Use the numeric keys to enter the provided value c Press ENTER to confirm your entry 3 The next value transducer delay represents the time the ultrasonic signal takes to travel fro...

Страница 162: ...ents the percent of peak the PT878GC uses to make measurements It will consider anything above the entered percentage as part of the signal To change this value a Press ENTER to open the text box b Us...

Страница 163: ...en a display window in Operate Mode To enter the option scroll to the Diagnostics entry on the Service Menu and press ENTER The screen appears similar to Figure 141 below Figure 141 The Diagnostics Wi...

Страница 164: ...or 20 mA Use the and arrow keys to move to the appropriate radio button and press ENTER 3 The next prompt asks you to enter the actual value shown on the multimeter or ammeter a Press ENTER to open t...

Страница 165: ...the input and set it to deliver the value entered in the previous step 5 Press the arrow key to move to the Read box and then press ENTER to read the low input 6 The next prompt asks for the value to...

Страница 166: ...Cutoff Velocity Averaging Errors Allowed Peak Detection Method Peak Thresholds Pulse and Code To enter the option scroll to the Signal Setup entry on the Service Menu and press ENTER The screen appea...

Страница 167: ...numeric keys to enter the desired value c Press ENTER to confirm your entry 5 The next prompt asks for velocity averaging the number of velocity measurements to average together to smooth out noise in...

Страница 168: ...dow appears similar to Figure 145 below Figure 145 The Measurement Mode Window in the Signal Setup Option The first prompt asks for the measurement mode detection method Use the and arrow keys to move...

Страница 169: ...do not select this option pressing the key returns the meter to the Transmit Code prompt 3 The next prompt asks how many pulses from 0 to 16 you wish to send a Press ENTER to open the drop down list...

Страница 170: ...rror message appears if the velocity falls outside these limits Repeat the steps in Step 1 on the previous page to enter the desired limits 3 The third prompt calls for the low and high limits for the...

Страница 171: ...ggles between English and metric measurements Thus the key will always display the opposite measurement from that which is currently active Press F1 to display the measurement in the alternate format...

Страница 172: ...Test entry on the Service Menu and press ENTER The screen appears similar to Figure 148 below Figure 148 The Test Option in the Service Menu Scroll to the desired entry in the option menu and press E...

Страница 173: ...To Continue Once you press a key a screen with a checkerboard pattern appears as shown in Figure 149 below Figure 149 A Functioning Test Screen Pressing a key two more times should result in two more...

Страница 174: ...n on the Test Menu and press ENTER The screen appears similar to Figure 150 below Press any key on the keypad and a window representing that key should darken Figure 150 The Test Keys Window Pressing...

Страница 175: ...51 below Figure 151 The Watchdog Test Screen Press F2 No to cancel the test and return to the Menu screen or press F3 Yes to start the test The PT878GC should go blank for a few seconds and then resta...

Страница 176: ...Use the numeric keys to enter the desired percentage c Press ENTER to confirm your entry You have completed entering data for the Impulse Response option Press F1 to start transmitting in the chosen...

Страница 177: ...ttery 1 If you wish to open the Part Number window press ENTER You can choose from a 1 8 Ahr NiCd battery part number 200 058 or a 3 0 Ahr NiMH battery part number 200 081 Scroll to the battery type y...

Страница 178: ...r otherwise record all logs settings and other data you wish to save IMPORTANT When uploading new software the PT878GC uses a bootloader program Be sure that the software you are installing is compati...

Страница 179: ...dure Be sure the infrared sensor on the meter has a clear path to the infrared adapter on the PC Now return to the PC 6 You now have two options In Windows Explorer browse to the file you wish to send...

Страница 180: ...ries Hyperterminal to open the Hyperterminal window 2 If the call is not connected click on Connect from the Call menu in Hyperterminal 3 From the File menu click Properties Be sure the New Connection...

Страница 181: ...low Figure 159 Flash Reloading Window 3 The meter now asks if you wish to reload the flash memory via the infrared communications port Press F3 Yes or C on the PC keyboard to begin reloading via IrCOM...

Страница 182: ...he Transfer Menu 2 Click on the replacement software designated by a cod extension from the folder where it has been stored 3 From the Protocol drop down menu select Xmodem 4 Click Send For a successf...

Страница 183: ...oblem The TransPort also has a set of diagnostic parameters to help you find and remedy the problem In general troubleshooting may be necessary if the TransPort displays an error message the flow read...

Страница 184: ...ror When they appear in a spreadsheet format errors are indicated by their number in the last row called Error The Error Code messages are only general descriptions of the possible problems Use Table...

Страница 185: ...s described in Chapter 9 on page 9 32 See Flowcell Problems on page 10 9 and Transducer Problems on page 10 11 E4 Signal Quality If too high electronic failure If too low flowcell electrical problem C...

Страница 186: ...signal in micro seconds N A N A Gain Up Displays upstream gain in dB N A N A Gain Dn Displays downstream gain in dB N A N A CNTup Displays the AGC DAC count for upstream gain setting N A N A CNTdn Dis...

Страница 187: ...Displays the normalization factor 0 85 1 0 0 85 Threshold Up Displays the Percent of Peak for the upstream transducer Set to 40 by default N A N A Threshold Dn Displays the Percent of Peak for the dow...

Страница 188: ...easurements 3 THE GAS SOUND SPEED MUST NOT VARY EXCESSIVELY The TransPort will tolerate relatively large changes in gas sound speed due to changes in gas composition and temperature However such chang...

Страница 189: ...g Any transducer damaged in this manner must be replaced Contact Panametrics Sensing for information on transducers in materials suitable for the application 3 INTERNAL DAMAGE An ultrasonic transducer...

Страница 190: ...Chapter 10 180 TransPort PT878GC User s Manual...

Страница 191: ...User s Manual 181 Chapter 11 Chapter 11 11 1 Specifications This section contains specifications for the following Operations and Performance Electronics Clamp on Transducers Thickness Gauge Option A...

Страница 192: ...to 0 5 of reading may be achievable with process calibration 11 1 1 6 Repeatability 0 2 to 0 5 of reading 11 1 1 7 Range Bidirectional See Table See Table 4 Table 5 and Table 6 11 1 1 8 Rangeability...

Страница 193: ...storage temperature exceeding 35 C 95 F is not recommended for more than one month 11 1 2 11 Standard Inputs Outputs One 0 4 to 20 mA current output One user selectable pulse solid state 5 V maximum o...

Страница 194: ...gas transducer clamping fixtures in polycarbonate carrying cases PCFG 1 for pipes smaller than 12 in 300 mm in diameter PCFG 2 for pipes larger than 12 in 300 mm in diameter PCFG 3 for pipes between...

Страница 195: ...ration to 260 C 500 F for 10 sec followed by 2 min air cooling 11 1 5 Additional Options 11 1 5 1 PC Communication The PT878GC communicates with a PC through the infrared interface and Windows operati...

Страница 196: ...Chapter 11 186 TransPort PT878GC User s Manual...

Страница 197: ...TransPort PT878GC User s Manual 187 Appendix A Appendix A A 1 Menu Maps Figure 162 The Main Menu...

Страница 198: ...Appendix A 188 TransPort PT878GC User s Manual...

Страница 199: ...Appendix A TransPort PT878GC User s Manual 189 Figure 163 The Site Menu...

Страница 200: ...Appendix A TransPort PT878GC User s Manual 190 Figure 164 The Program Menu Transducer Lining Fluid and Path Options...

Страница 201: ...Appendix A TransPort PT878GC User s Manual 191 Figure 165 The Program Menu Pipe Option...

Страница 202: ...Appendix A TransPort PT878GC User s Manual 192 Figure 166 The Program Menu Scan Measure Analog Input and Analog Output Options...

Страница 203: ...Appendix A TransPort PT878GC User s Manual 193 Figure 167 The Program Menu Digital Output User Function and Correction Factor Options...

Страница 204: ...Appendix A TransPort PT878GC User s Manual 194 Figure 168 The Meter Menu...

Страница 205: ...Appendix A TransPort PT878GC User s Manual 195 Figure 1 Logging Mneu...

Страница 206: ...Appendix A TransPort PT878GC User s Manual 196 Figure 169 The Logging Menu Figure 170 The Service Menu Figure 1 Service Menu...

Страница 207: ...various past and current Microsoft Windows operating systems IrCOMM is an infrared standard supported by Microsoft while OBEX is a newer infrared standard which allows more straightforward use and fas...

Страница 208: ...Windows Me 98SE 98 95 The Windows 98SE and Me operating systems include IR drivers However in most cases the IR port has not been installed or configured From the Control Panel click on Add New Hardw...

Страница 209: ...rush or file before measurements are attempted Generally it is possible to make corrosion measurements through thin layers of rust as long as the rust is smooth and well bonded to the metal below Some...

Страница 210: ...and care with which the gauge has been calibrated It is essential that the thickness gauge be calibrated as described on page 9 12 whenever the transducer is changed or you have a reason to doubt the...

Страница 211: ...ata Sheets for Couplants A variety of couplants are available for use with Panametrics Sensing transducers To obtain a material safety data sheet MSDS for any couplant Panametrics Sensing supplies go...

Страница 212: ...Appendix D 202 TransPort PT878GC User s Manual...

Страница 213: ...Clamp On Transducers Path Parameters for 50 Clamp on Transducers Installing 15 Clearing Softkey 97 Communications Option 114 Configuring the display 89 Connections 5 Analog Inputs 7 Analog Outputs 7 P...

Страница 214: ...Interface IrCOMM 168 IrOBEX 168 Infrared Wireless Interface 7 Initial Setup 5 Input Output 3 Input Output Connections 7 Inputs Analog Inputs 7 Analog Entering 61 in Standard Volume Option 52 Inputs O...

Страница 215: ...e Lining Entering 47 Pipe Material 45 Pipe Materials 182 for Thickness Gauge 185 Pipe Problems 178 Pipe Sizes 182 Pipe Thickness Range for Thickness Gauge 185 Pipe Wall Thickness 182 Power Connections...

Страница 216: ...Air 182 Pressure Requirements Other Gases 182 Range 182 Rangeability 182 Repeatability 182 Thickness Gauge 185 Transducers 184 Standard Volume Option Entering Inputs 52 Entering Parameters 50 Static...

Страница 217: ...0 Troubleshooting 173 Error Messages 174 Turning On Off 9 Tw time delay 43 U Units Option 106 User Functions Option 68 User Functions Entering 68 User Table Option 117 User Tables Creating 117 V V4 an...

Страница 218: ...Index 208 TransPort PT878GC User s Manual...

Страница 219: ...EN 61326 1998 Class A Annex C Continuous Unmonitored Operation following the provisions of the 89 336 EEC EMC Directive The units listed above and any transducers supplied with them spoolpieces are ad...

Страница 220: ...2 Doc Tittle Manual Type...

Страница 221: ...r purpose and warranties arising from course of dealing or usage or trade Return Policy If a Panametrics Sensing instrument malfunctions within the warranty period the following procedure must be comp...

Страница 222: ...Warranty 4 TransPort PT878GC User s Manual no content intended for this page...

Страница 223: ......

Страница 224: ...and company names are trademarks of their respective holders 910 229 Rev B3 Customer Support Centers U S A The Boston Center 1100 Technology Park Drive Billerica MA 01821 U S A Tel 800 833 9438 toll...

Отзывы: