Chapter 8.
126
TransPort™ PT878GC User’s Manual
Figure 117: The General Log Format Window
To step through each parameter, press the [
] key.
1.
The first prompt asks you to choose between a linear or circular format for the log. (A linear log stops when it
reaches a specified end time or the PT878GC runs out of memory, while a circular log runs continuously until
manually stopped.)
a.
Use the [
] and [
] keys to scroll to the appropriate radio button.
b.
Press
[ENTER]
to confirm your selection.
2.
The next prompt asks you to choose whether the log is to be a standard log, recording specified data, or an error
log (one which records error conditions).
a.
Use the [
] and [
] keys to scroll to the appropriate radio button.
b.
Press
[ENTER]
to confirm your selection.
3.
The next prompt asks for the starting date and time.
a.
Press
[ENTER]
to enter the text box. The meter highlights the first number. Use the [
] and [
] keys to scroll to any
number you wish to change, or the [
] key to scroll to the time box.
b.
Two alternatives are available to change a highlighted number:
• Use the numeric keys to enter the desired number.
• Use the [
] or [
] arrow keys to scroll, in 1-digit increments, to the desired number. (For example, if the text
box displays 09, pressing the [
] key twice changes the number to 07.) You can scroll from 01 to 12 for the
month and from 01 to 31 for the day (depending on the number you have selected for the month).
c.
In either case, press
[ENTER]
to confirm the entry.
4.
Follow the same procedure to enter the end date and time.
5.
The final prompt in this window asks for the logging interval.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired interval in seconds.
c.
Press
[ENTER]
to confirm the entry.
6.
You have completed entering parameters in this window. To select the measurements,
a.
Press the [
] key to return to the Format prompt.
b.
Then press the [
] key to return to the General tab.
General
Measurements
Cancel
Activate
Log Name 10SEC.LOG
End Date/Time
Standard
Error
Linear
Circular
Type
Format
Logging Interval
Start Date/Time
01:38:08
2004/11/01
10
09:38:08
secs
2004/11/01
Содержание PT878GC
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii TransPort PT878GC User s Manual...
Страница 16: ...Chapter 2 6 TransPort PT878GC User s Manual Figure 3 Connection Locations...
Страница 50: ...Chapter 3 40 TransPort PT878GC User s Manual...
Страница 98: ...Chapter 5 88 TransPort PT878GC User s Manual...
Страница 148: ...Chapter 8 138 TransPort PT878GC User s Manual...
Страница 152: ...Chapter 9 142 TransPort PT878GC User s Manual Figure 131 Printout of a Typical Drive Report...
Страница 190: ...Chapter 10 180 TransPort PT878GC User s Manual...
Страница 196: ...Chapter 11 186 TransPort PT878GC User s Manual...
Страница 197: ...TransPort PT878GC User s Manual 187 Appendix A Appendix A A 1 Menu Maps Figure 162 The Main Menu...
Страница 198: ...Appendix A 188 TransPort PT878GC User s Manual...
Страница 199: ...Appendix A TransPort PT878GC User s Manual 189 Figure 163 The Site Menu...
Страница 200: ...Appendix A TransPort PT878GC User s Manual 190 Figure 164 The Program Menu Transducer Lining Fluid and Path Options...
Страница 201: ...Appendix A TransPort PT878GC User s Manual 191 Figure 165 The Program Menu Pipe Option...
Страница 204: ...Appendix A TransPort PT878GC User s Manual 194 Figure 168 The Meter Menu...
Страница 205: ...Appendix A TransPort PT878GC User s Manual 195 Figure 1 Logging Mneu...
Страница 212: ...Appendix D 202 TransPort PT878GC User s Manual...
Страница 218: ...Index 208 TransPort PT878GC User s Manual...
Страница 220: ...2 Doc Tittle Manual Type...
Страница 222: ...Warranty 4 TransPort PT878GC User s Manual no content intended for this page...
Страница 223: ......