Chapter 9.
156
TransPort™ PT878GC User’s Manual
b.
If you are satisfied with the calibration, press the [
] key to return to the Input# prompt, and repeat steps 2
through 8 for the other input.
You have completed calibrating the inputs. To leave the Inputs window,
• Press the [
] key to return to the Inputs tab, and the [
] arrow key to move to the Outputs tab.
• Press
[F2]
(Cancel) to return to Operate Mode without confirming the changes.
• Press
[F3]
(OK) to confirm the new data. The PT878GC returns to Operate Mode.
9.1.8 Setting up Signal Parameters
The Signal Setup option enables you to set parameters that affect the transducer signal:
• Delta-T Offset
• Transmit Sample Size
• Zero Cutoff
• Velocity Averaging
• Errors Allowed
• Peak Detection Method
• Peak Thresholds
• Pulse and Code
To enter the option, scroll to the Signal Setup entry on the Service Menu and press
[ENTER]
. The screen appears
similar to Figure 144 below.
Figure 144: The Signal Parameter Window
1.
Press the [
] arrow key to enter the window.
2.
The first prompt asks for the Delta-T offset. Delta-T is the difference between the upstream and downstream
transit times of the transducers. The Delta-T offset should normally be set to zero.
Note:
Consult Panametrics Sensing before performing this step.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the recommended value.
c.
Press
[ENTER]
to confirm your entry.
Meas Mode
Signal Param
Signal Setup
Cancel
OK
Delta-T Offset
Errors Allowed
Velocity Averaging
Transmit Sample Size
Zero Cutoff
0 m/s
10
8
5 ns
8
Pulse/Code
Содержание PT878GC
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents viii TransPort PT878GC User s Manual...
Страница 16: ...Chapter 2 6 TransPort PT878GC User s Manual Figure 3 Connection Locations...
Страница 50: ...Chapter 3 40 TransPort PT878GC User s Manual...
Страница 98: ...Chapter 5 88 TransPort PT878GC User s Manual...
Страница 148: ...Chapter 8 138 TransPort PT878GC User s Manual...
Страница 152: ...Chapter 9 142 TransPort PT878GC User s Manual Figure 131 Printout of a Typical Drive Report...
Страница 190: ...Chapter 10 180 TransPort PT878GC User s Manual...
Страница 196: ...Chapter 11 186 TransPort PT878GC User s Manual...
Страница 197: ...TransPort PT878GC User s Manual 187 Appendix A Appendix A A 1 Menu Maps Figure 162 The Main Menu...
Страница 198: ...Appendix A 188 TransPort PT878GC User s Manual...
Страница 199: ...Appendix A TransPort PT878GC User s Manual 189 Figure 163 The Site Menu...
Страница 200: ...Appendix A TransPort PT878GC User s Manual 190 Figure 164 The Program Menu Transducer Lining Fluid and Path Options...
Страница 201: ...Appendix A TransPort PT878GC User s Manual 191 Figure 165 The Program Menu Pipe Option...
Страница 204: ...Appendix A TransPort PT878GC User s Manual 194 Figure 168 The Meter Menu...
Страница 205: ...Appendix A TransPort PT878GC User s Manual 195 Figure 1 Logging Mneu...
Страница 212: ...Appendix D 202 TransPort PT878GC User s Manual...
Страница 218: ...Index 208 TransPort PT878GC User s Manual...
Страница 220: ...2 Doc Tittle Manual Type...
Страница 222: ...Warranty 4 TransPort PT878GC User s Manual no content intended for this page...
Страница 223: ......