DigitalFlow™ GC868 Programming Manual (2-Channel)
1-18
Chapter 1. Programming Site Data
1.6.5.4
Setting Advanced
Parameters
In the
ADVAN
option, you can enter parameters for a variety of specialized
operations. You can set parameters for static density, program tables for steam or
natural gas and for multiple K factors, and modify parameters for the Skan and
Skan/Measure techniques. To enter the
ADVAN
submenu, press
[F4]
at the
SETUP
window, and refer to Figure A-3 on page A-3.
1.6.5.4.1 Static Density
Use this option to specify basic parameters for mass flow measurement.
1.
To enter the
MASS
submenu, press
[F1]
at the
ADVAN
window.
2.
At the
Static Density?
prompt, press
[F1]
=
NO
or
[F2]
=
YES
. If you press
NO
, the
GC868 returns to the
ADVAN
window.
3.
If you press
YES
, select the
Density Type
to calculate Mass Flow
:
[F1]
=
Rho
(Fluid
Density) or
[F2]
=
Mw (Molecular Weight)
.
4.
The next prompt depends upon your selection in step 3.
•
If you select Rho
,
a.
The GC868 asks for the type of volumetric flow. Press
[F1]
to select
AcVOL
(Actual Volumetric Flow) or [F2] to select StVOL (Standard Volumetric
Flow).
b.
Then enter the
Fluid Density
of the gas to be measured and press
[ENT]
. (The
meter will only accept values from 0.00001-100 lb/ft
3
.)
• If you select Mw,
Enter the Molecular Weight of the gas to be measured and press
[ENT]
. (The
range is from 0 to 100 glmol, and the default value is 28.85 glmol.)
The GC868 program returns to the
ADVAN
window.
1.6.5.4.2 Setting the A Value
In the Skan/Measure Integrate mode, the A value aids in minimizing cycle skipping.
Use this option to specify parameters for the A value.
IMPORTANT:
The
AVAL
default settings are suitable for most applications. Before
changing any of these parameters, consult the factory.
1.
To enter the
AVAL
submenu, press
[F2]
at the
ADVAN
window.
2.
Use the numeric keys to enter the desired
A_Value Minimum
, and press
[ENT]
.
(The available range is from 75 to 100.)
3.
Use the numeric keys to enter the desired
A_Value Maximum
, and press
[ENT]
.
(The available range is from 100 to 125.)
4.
Enter the desired
-Offset % of Peak
from 0 to 95%, and press
[ENT]
. (The default is
20.)
5.
Enter the number of
Cycle Count Samples
(from 1 to 40) and press
[ENT]
. (The
default is 8.)
6.
Enter the
Cycle #
(from 1 to 100) and press
[ENT]
. (The default is 1.)
The GC868 returns to the
ADVAN
window.
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 78: ...Chapter 5 Clearing Data 5 3 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 90: ...Appendix A Menu Maps A 1 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 130: ...Index 4 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...