Chapter 1. Programming Site Data
1-17
DigitalFlow™ GC868 Programming Manual (2-Channel)
After completing these steps, the program returns to the
SETUP window.
Table 1-4 on
the next page lists the default values and limits for SETUP parameters.
1.6.5.2
Setting Response
Time
Use this option to specify the number of readings that occur before the meter will
respond to a step change in flow rate. In general, the smaller the number of
readings, the less steady the display will appear.
1.
Press
[F2] = AVRG to enter the AVERAGE submenu.
2.
Use the
[
→
]
and
[F1]-[F4]
keys to select the desired option. The available
response time options are
1, 2, 5, 10, 30, 60 readings and STATS
. For best results,
select
STATS
(statistics), as this increases the response time under steady flow
conditions while still allowing a rapid response to changes in flow rate.
After completing this step, the program returns to the
SETUP
window.
1.6.5.3
Initializing the
System
Use this option to initialize (reset) all of the parameters within the
SETUP
menu to
their default values.
1.
Press
[F3] = INIT.
2.
Press
[F1] = NO
to keep the current values or press
[F2] = YES
to reset all values
to their default settings.
Table 1-4: Default Values and Limits for SETUP Parameters
Parameter
Default
Value
Low Limit
High Limit
Signal Low Limit
50
-20
100
Signal High Limit
68
25
100
Cor. Peak Limit
400
0
500
Soundspeed ±
Limit
10%
1%
50%
Velocity Low Limit
-75.0 ft/sec
(-23 m/sec)
-500 ft/sec
(-150 m/sec)
+500 ft/sec
(+150 m/sec)
Velocity High Limit
75.0 ft/sec
(23 m/sec)
-500 ft/sec
(-150 m/sec)
+500 ft/sec
(+150 m/sec)
Acceleration Limit
5.0 ft/sec
2
(4.6 m/sec
2
)
0
100 ft/sec
2
(30 m/sec
2
)
Amp. Discrim Low
14
0
100
Amp. Discrim High
34
0
100
Delta T Offset
0
-1,000 µsec
1,000 µsec
Skan T Offset
12.6 µsec
-500 µsec
500 µsec
% of Peak
80%
1%
100%
XMIT Sample Size
8
2
32
M>S Switch
5 µsec
0 µsec
250 µsec
# shifts
3
0
10
A Divisor
2.5
0.1
10
# Transmit Pulses
1
1
16
T Window
1
0
1,000
R Window
10
10
128
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 78: ...Chapter 5 Clearing Data 5 3 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 90: ...Appendix A Menu Maps A 1 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 130: ...Index 4 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...