Chapter 1. Programming Site Data
1-11
DigitalFlow™ GC868 Programming Manual (2-Channel)
f.
Use the numeric keys to enter the
Wedge Angle
(the angle of the ultrasonic
transmission) provided by the factory in degrees and press
[ENT]
.
• If you selected
OTHER
in step d, use the numeric keys to enter the
Wedge
Soundspeed
in ft/sec or m/sec provided by the factory and press
[ENT]
. The
meter proceeds to step 5.
1.6.3.2
Wedge
Temperature
4.
Use the numeric keys to enter the
Wedge Temperature
in deg F provided by the
factory and press
[ENT]
.
1.6.3.3
Pipe Material
5.
Press
[F1]-[F4]
to select the
Pipe Material
. Press
[
→
]
to access additional
options, as listed in Table 1-3 below.
• Depending on the pipe material choice, another prompt may ask you to select
the
Specific Material
.
• If you have selected
OTHER
, enter the pipe material
Sound Speed.
Press
[ENT].
1.6.3.4
Pipe OD
6.
Enter the known
Pipe OD
or circumference and use the
[F1]-[F4]
keys to select
the appropriate units. Press
[ENT].
(The meter will only accept values from
1/8 to 648 in.)
The required pipe data may be obtained by measuring either the pipe outside
diameter (OD) or circumference at the transducer installation site. Alternatively, the
information may be obtained from tabulated data for standard pipe sizes.
1.6.3.5
Pipe Wall
7.
Enter the known thickness of the
Pipe Wall
, in the same units used for the pipe
OD. Press
[ENT]
. (The meter will only accept values from 0 to 4 in.)
If the pipe wall thickness is not known and cannot be conveniently measured, look
up the value in a table of standard pipe size data or use the Model GC868’s on-line
Help Menu
.
1.6.3.6
Fluid Type
8.
Press
[F1]
= natural gas,
[F2]
= air,
[F3]
= steam, or
[F4]
= any other gas to
select the
Fluid Type.
• If
OTHER
was selected, enter the speed of sound (in feet per second) in the gas
to be measured. Press
[ENT]
. (The meter will only accept values from
125 to 9,000 ft/sec.)
1.6.3.7
Reynolds
Correction
9.
The
Reynolds Correction
factor is a number based on the kinematic viscosity
and flow rate of the gas. At the
prompt, press
[F1] to turn Reynolds Correction
off,
or
[F2] to turn it on.
• If you select Reynolds Correction, the GC868 asks for the
Kinematic Viscosity
of
the gas (available in the brochure
Sound Speeds and Pipe Size Data
, 914-004).
Use the numeric keys to enter the correct value and press
[ENT]
.
1.6.3.8
Calibration Factor
10.
Enter a value for the flow
Calibration Factor
and press
[ENT]
. The default value is
1.00. (The meter will only accept values from 0.5000 to 2.0000.)
Table 1-3: Pipe Materials
Pipe Material Category
Specific Material
Steel
Carbon Steel or Stainless Steel
Iron
Ductile Iron or Cast Iron
Cu - Copper
None
Al - Aluminum
None
Brass
None
CuNi - Copper/Nickel
70% Cu 30% Ni or 90% Cu 10% Ni
Glass
Pyrex, Flint, or Crown
Plastic
Nylon, Polyester, Polypropylene, PVC
(CPVC), Acrylic
Other*
Any material
Содержание DigitalFlow GC868
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 6: ...Warranty iv DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 10: ...Contents viii DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 78: ...Chapter 5 Clearing Data 5 3 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 90: ...Appendix A Menu Maps A 1 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...
Страница 130: ...Index 4 DigitalFlow GC868 Programming Manual 2 Channel no content intended for this page...