3. Fitment of Bottom Manifold
1. Position the manifold on base plate with the near tab
below the cross pin and the side tabs against the
support legs.
2. Push the manifold down flat onto the base plate.
3. Ensure that the gasket is flat within the retaining pegs of
the manifold.
4. Placement of First Filter Capsule
1. Center the capsule slightly above the manifold with the
side tabs latched around the support legs. Push the
capsule down to seat it on the bottom manifold.
2 Check the position of the gasket. It must lie flat fully
within the capsule recess.
5. Placement of Additional Filter Capsules
1. Place second capsule into position slightly above the
first capsule. Securely latch the side tabs around the
support legs.
2. Push the capsule down flat.
3. Check the position of the gasket. It must lie flat fully
within the recess.
6. Assembly of Changing Plate
1. Position the manifold on the top capsule.
2. Place clamping plate on top of the vent manifold with
rear support positioned between guide pins.
9
USD 2554c
www.pall.com/biopharm
3. Unscrew fully the wing bolts and wind spindle down
through hole in retaining plate.
4. Slide retaining plate into spindle groove.
5. Align clamping plate with the two screw holes and
apply wing bolts and tighten securely.
7. Bottom In/Bottom Out Flow Version
1. Rotate torque wheel to clamp down capsules until
torque limiter clicks out and wheel rotates freely.
Check to ensure that capsules are fully clamped down.
2. Connect inlet and outlet hoses on the manifold
as indicated.
3. Open vent valves
4. Begin filling capsule until fluid appears.
5. Close valves.
6. Check for leaks.