8. Bottom In/Top Out Flow Version
1. Position top manifold with gasket facing down
onto capsule.
2. Attach hose support bracket securely to clamping plate.
3. Rotate torque wheel clockwise to clamp down capsules
until torque limiter clicks out and wheel rotates freely.
4. Connect inlet line to bottom manifold.
5. Connect outlet line to top manifold.
6. Cover open manifold ports with appropriate control
valves. A control valve should be fitted to the top inlet
pipe to assist with purging air and post use blow down.
12
USD 2554c
Follow the instructions for the Pilot Scale Chassis in Appendix
A: Chassis User Instructions — Pilot Scale on page 8, except:
1. Place a vent manifold on top of the first media grade capsule(s).
2 Place a distribution manifold on top of the first vent manifold.
3. Place the capsules of the second media grade on top of
the second distribution manifold.
4. Place a vent manifold on top of the second media
grade capsule(s).
5. Clamp capsules and manifolds using the torque wheel.
6. Connect inlet and outlet lines to the distribution manifolds as
indicated by the direction of flow arrows on the manifolds.
7. Open vent valves and commence filling.
8. Close vent valves when fluid appears and check for leaks.
Appendix C: Chassis User Instructions — Series Filtration
Pilot Scale Chassis with Series Filtration Setup
For Series Flow filtration using two different media grades, the chassis will accommodate
combinations of large, medium and small capsules. Please contact your local Pall office or
distributor for guidance on combinations of capsule sizes.
Process Scale Chassis with Series Filtration Setup
For Series Flow using capsules of two different media grades, the chassis will accommodate
up to 9 large capsules in the 10-high chassis and up to 4 large capsules in the 5-high chassis.
Follow the instructions for the Process Scale Chassis in Appendix B: Chassis User Instructions
— on page 10, except:
1. Place a vent manifold on top of the first media grade capsule set
2. Place a distribution manifold on top of the first vent manifold
3. Place the capsule of the second media grade set on top of
the second distribution manifold
4. Place a vent manifold on top of the second media capsule set.
5. Clamp capsules and manifolds using the torque wheel.
6. Connect inlet and outlet lines to the distribution manifolds
as indicated by the direction of the flow arrows on the
manifolds.
7. Open vent valves and commence filling.
8. Close vent valves when fluid appears and check for leaks.