1. Introduction
The following procedures must be followed for the installation of the Pall Stax filter system.
The instructions contained in the product documentation must be read thoroughly as they
contain valuable information gained by extensive experience. It is very important that all
instructions are carefully followed and where appropriate they should be incorporated into the
end user’s standard operating procedures. If some of the procedures do not suit your needs,
please consult Pall or your local distributor before finalizing your system.
Use of this product in a manner other than in accordance with Pall’s current recommendations
may lead to injury or loss. Pall cannot accept liability for such injury or loss.
2. Specifications
The maximum working pressures and temperatures are specified in Table 1. Operation outside
the specifications and with fluids incompatible with the materials of construction may cause
personal injury and result in damage to the equipment. Incompatible fluids are fluids which
chemically attack, soften, stress, attack or adversely affect the materials of construction.
Please refer to Pall for a list of incompatible fluids.
Table 1
Operating Conditions
Maximum Operating Pressure
3.5 bar at 25 °C
1.0 bar at 60 °C
Maximum Operating Temperature
25 °C at 3.5 bar
60 °C at 1.0 bar
Recommended Operating Conditions Maximum differential Pressure 2.4 bar at 3.5 bar operating pressure
Contact Pall for advice in case of prolonged exposure at extremes of operating temperatures.
Warning: European Directive 94/9/EC (ATEX) Equipment Intended for Use in Potentially
Explosive Atmospheres For information relating to European Directive 94/9/EC (ATEX),
please refer to page 7. For information relating to Zone 0/20 Applications, please
contact Pall. More information can be obtained through Pall, your local distributor or the
Pall website.
3. Receipt of Equipment
Caution: Stax standard capsule and manifold assemblies are suitable for sterilization by
autoclaving. Stax carbon capsules incorporating Seitz
®
AKS media must not be
autoclaved. Please check the product label prior to use to ensure product part numbers
correspond to the application. Stax capsules and manifolds are supplied non-sterilized,
if unsure of suitable sterilization method, please contact Pall.
1. Store the capsules and manifold assemblies in clean, dry conditions between 0 °C and
30 °C without exposure to irradiation sources such as direct sunlight, and wherever
practical in the packaging as delivered.
2. Do not remove from packaging until just before installation.
3. Check that the bag or packaging is undamaged prior to use.
4. Ensure that the type of capsule assembly selected is suitable for the application.
5. In addition to the part number, each capsule assembly is identified by a unique identification
batch number and a unique serial number.
6. Stax capsule modules should be used within 36 months after their production date as
shown on the box label with the exception of Stax capsules with Bio-Series media, which
must be used within 18 months after their production date.
7. The manifold kits are equipped with protective caps on the inlet and oulet connections.
Please ensure these caps are removed prior to putting your system into operation.
2
USD 2554c
!
!
216495 - Pall Europe Ltd GN12.5291.qxd:Layout 1 1/6/12 14:10 Page 2