background image

Reset

NO

. Exits from reset menu options without action.

FACTORY RESET

. Returns all settings except

date/time and ATC calibration to factory default
values after 

ENTER

is pressed then restarts meter.

DATA RESET

. Erases data stored in memory while

retaining other settings after 

ENTER

is pressed.

CALIBRATION RESET

. Erases non-ATC

calibration data while retaining other settings
after 

ENTER

is pressed.

pH Calibration

For best results, periodic calibration with known,
accurate standards is recommended. Calibrate with
standards that bracket your intended measuring
range while including a neutral standard (pH 7.00
or 6.86). For example, if you expect to measure
samples from pH 6.2 to pH 9.5, calibration with
4.01, 7.00, and 10.01 standards will work well.
Provide stirring for best results. After calibration with
two or more points, the active slope segment of the
measurement will be visible on the bottom display
during measurement. 100% slope will be shown if
only one calibration point is performed and “- - -“
if no calibration is performed. The meter will
automatically return to measurement mode upon
successful completion of the number of specified
calibration points. To specify a different number
of pH calibration points, see 

pH Buffer Options

.

Using Automatic Buffer Recognition

1

While in pH measurement mode, dip the pH
and ATC sensor(s) into your first standard, then
press 

CAL

. The primary display will search for

the nearest standard value, while the secondary
display will show the un-adjusted value.

2

When the 

“READY”

indicator appears, press

ENTER

to accept. The primary reading will

flash 

“DONE”

.

3

Rinse your electrode(s) then dip into the next pH
standard. The primary display will search for the
nearest standard value that has not yet been
calibrated, while the secondary display will show
the unadjusted value. When the 

“READY”

indicator appears, press 

ENTER

to accept.

4

To calibrate another pH standard repeat Step 3 or
press 

MEAS

to return to pH measurement mode.

Using Manual Recognition /
Custom Buffers

1

While in pH measurement mode, dip the
pH and ATC sensor(s) into your first standard
then press 

CAL

.

2

When the 

READY

indicator appears, use

up/down arrows to adjust the primary reading
to match the standard value at the measured
temperature, then press 

ENTER

.

3

Rinse your electrode(s) then repeat Step 2
with a standard that is 

1 pH unit from the

previous standard value.

4

To calibrate another pH standard, repeat
Step 3 or press 

MEAS

to return to pH

measurement mode.

mV Offset Adjustment

1

While in mV measurement mode, dip the ORP
and ATC sensors into a solution with a known
mV value (i.e. Zobel, Light’s, quinhydrone, or
iodide/triiodide) and stir.

2

When the 

“READY”

indicator appears, use

up/down arrows to adjust the primary reading
to match the mV value at the measured
temperature, then press 

ENTER

. The meter

allows an adjustable maximum value of ±200
mV from the factory default mV value. When
an offset has been stored successfully, R.mV
replaces mV.

Temperature Calibration/
Manual ATC

1

Press 

CAL

from any measurement,

then press 

MODE

.

2

Skip to step 3 for manual ATC, otherwise, dip
the temperature sensor into a solution with a
known accurate temperature. The upper display
shows the active temperature while the lower
display shows the factory default temperature
without adjustment.

3

Use up/down arrows to adjust the upper
display. Press 

ENTER

to accept the calibration

temperature. The maximum adjustable value
is ±10ºC (or ±18ºF) from factory default.

3

Micro 800 pH mV Operating Instructions

Содержание Micro 800

Страница 1: ...Palintest ZI INST 1330 Palintest Micro 800 pH mV Operating Instructions...

Страница 2: ...R to view mV Offset and Slope of the measured reading System Settings Data Logging MANUAL upon key press only TIMED interval Choose SEC MIN HOUR interval Automatic shut off after 10 minutes Choose ON...

Страница 3: ...dard The primary display will search for the nearest standard value that has not yet been calibrated while the secondary display will show the unadjusted value When the READY indicator appears press E...

Страница 4: ...rument Do not use any aggressive cleaning chemicals solvents or similar agents There are no user serviceable parts inside Attempts to service internal parts may void the warranty WARNING No modificati...

Страница 5: ...ope del valor medido Ajustes del Sistema Registro de datos MANUAL al apretar un bot ns lo intervalo TIMED Cronometrado Elija SEG MIN HORA intervalo Apagado autom tico despu s de 10 minutos Elija ON od...

Страница 6: ...iguiente est ndar de pH La visualizaci n primaria va a buscar al valor est ndar m s cercano que no ha sido calibrado a n mientras que la visualizaci n secundaria va a mostrar el valor no ajustado Cuan...

Страница 7: ...esivos solventes o substancias similares No hay partes en el interior que pueden ser mantenidos por el usuario El intento de mantener partes internas puede anular la garant a ADVERTENCIA No est permit...

Страница 8: ...SEC MIN HOUR Sec min heure Arr t automatique apr s 10 minutes S lectionner ON ou OFF R glages de l horloge Date S lectionner le syst me USA am ricain MM JJ AAAA ou Euro europ en JJ MM AAAA Dur e Choi...

Страница 9: ...l indique DONE Effectu 3 Rincer l lectrode ou les lectrodes puis la tremper dans votre prochain talon pH L cran principal recherchera la valeur d talon la plus proche qui n a pas encore t talonn e tan...

Страница 10: ...dans l appareil Ne pas utiliser de produits chimiques de nettoyage agressifs solvants ou produits similaires L appareil ne contient aucune pi ce r parable par l utilisateur Toute tentative de r parati...

Страница 11: ...und Steigung des Messwertes zu sehen Systemeinstellungen Messwerterfassung MANUELL bei Bet tigen der Taste nur TIMED Zeitlich Festgelegtes Intervall W hlen Sie ein Intervall SEK MIN STUNDE Automatisc...

Страница 12: ...in den n chsten pH Standard Das prim re Display wird nach dem n chstgelegenen Standardwert der noch nicht kalibriert wurde suchen w hrend die sekund re Anzeige den unangepassten Wert anzeigen wird So...

Страница 13: ...Verwenden Sie keine aggressiven Reinigungschemikalien L sungsmittel oder hnliche Substanzen Das Ger t enth lt keine vom Benutzer zu wartende Teile Versuche die Innenteile zu warten k nnte die Garanti...

Страница 14: ...00 131516 Fax 61 1300 131986 palintest palintest com au Palintest USA 1455 Jamike Avenue Suite 100 Erlanger Kentucky 41018 USA Tel 1 859 341 7423 Toll Free 800 835 9629 Fax 1 859 341 2106 info palinte...

Отзывы: