5.5. Manual Control / Calibration
Manual pump control
group box
Allows you to manually control the various pumps and their calibration.
Pump flow
There are three modes to choose from:
1. Fast flow
2. Moderate flow
3. Precise flow
Depending on the tubing used, you have the option to achieve the following performance:
a) high flow Pharmed/Viton/Santoprene
• Fast flow - maximum average flow 110ml/min
• Moderate flow - maximum average flow 60ml/min
• Precise flow - not suggested (in-regular flow depending from material hardness.
The best for precise dose is silicone tubing.
b) high flow special silicone tube (thick)
• Fast flow - maximum average flow 140ml/min
• Moderate flow - maximum average flow 110ml/min
• Precise flow - maximum average flow 50ml/min
c) precise flow special silicone tube (thin)
• Fast flow - maximum average flow 60ml/min
• Moderate flow - maximum average flow 26ml/min
• Precise flow - maximum average flow 10ml/min (!)