Page 43
Kore 7th Manual / kHLab Edition
Pump Calibration Using the Calibration/Measuring Cylinder
1) Connect tubes to the pump entrance and pump exit and then place the pump
entrance tube in the proportioned liquid.
2) Remove air from the tubes by manual activation of the pump.
3) Set up proportioning velocity (it is recommended to use
Fast Flow
for tubing
typically embedded in the pump heads).
4) Select the
calibration mode
.
5)
Select a number of the pump
that is to be calibrated.
6) Arrange a proportioning vessel, preferably a measuring cylinder.
7) Press
Start
button.
8) Upon completion, read accurate amount of the liquid in the measuring cylinder and
then enter it’s amount to the
Measured liquid amount field [in ml]
.
9) Save the calibration by pressing
Save
button.
10) Use
Manual dose
option to check the calibration correctness.
11) Should the amount of proportioned liquid differ from the value that has been
saved during the calibration check, it will be necessary to repeat calibration paying
attention to air bubbles in the tubes. The entire tubing shall be filled in with the
liquid.
12) Calibration shall be carried out for all the pumps individually and the results shall be
entered upon it's completion.
Pump Calibration
group box
Manual dose [ml]
Allows manual dispensing of a specified quantity of fluid.
Liquid low level alarm
group box
Allows you to set an alarm for a minimum level of dispensed liquids. You can also turn on/
off sound alarm for low level of liquid in bottles.
Magnetic Stirrer settings
group box
Stirrer model
V1.0 – it’s the first generation of stirrers without external power supply
(power taken from Kore 7th dosing station).
V2.0 – it’s the new, next generation stirrer, with own power supply.
Stirring time
Stirring time before starting dose. Minimum 5s – maximum 60s. We
suggest between 30 and 40 seconds (depend from liquid density/
type).
Stirring method
7 different stirring programs. Different speed and variable „pulsations”
mode. Help choose the best one for used bottle type and liquid
density. Test before the
save
. Magnetic pellet should spin without any
obstacles or bouncing.
Manual stirring
Allows to run manual stirring.
Allows for configure proper version of
Magnetic Stirrer
and set stirring time before doses.