RANGE COMMANDS
These commands select either a fixed range or AUTO range. See Section 3 for more information.
:SENSe:FLUX:RANGe:AUTO
Selects the AUTO RANGE function.
:SENSe:FLUX:RANGe <n>
Selects a fixed range <n>, where n =
0 for 300 G / 30 mT / 23.88 kA/m
1 for 3 kG / 300 mT / 238.8 kA/m
2 for 30 kG / 3 T / 2388 kA/m
:SENSe:FLUX:RANGe?
Returns an ASCII digit representing the present range setting for the meter, as follows:
0 for 300 G / 30 mT / 23.88 kA/m
1 for 3 kG / 300 mT / 238.8 kA/m
2 for 30 kG / 3 T / 2388 kA/m
4-18