RANGE SELECTION
The meter is capable of providing flux density measurements on one of three fixed ranges, or it can be
programmed to automatically select the best range for the present flux density. The available ranges are listed in
the
SPECIFICATIONS
section of this manual. The ranges advance in decade steps. The lowest range offers the
best resolution while the highest range allows higher flux levels to be measured.
In the autorange mode, the range is advanced if the reading reaches 95% of the present range. The range is
lowered when the reading falls below 9% of full scale for the present range. The speed at which the readings are
updated decreases slightly when AUTO ranging is used.
NOTE:
The AUTO range selection will be canceled if the RELATIVE mode is turned on.
To choose the desired range, press the RANGE pushbutton. The “RANGE” legend will flash. Press the UP (
5
) and
DOWN (
6
) pushbuttons to select the desired range. To choose AUTO RANGE on the display, press the SHIFT
pushbutton followed by the RANGE pushbutton.
This setting is saved and will be restored the next time the meter is turned on.
Figure 3-12
RANGE Function
3-7