ANALOG OUTPUT USAGE
See the SPECIFICATIONS section for frequency range and accuracy of the analog output.
The analog output signal is calibrated to 3VDC or 3VRMS depending on the selected mode. The analog output
voltage is proportional to the flux density being measured depending on the meters current range. For instance, when
in the 30mT DC range a reading of 12.3mT relates to an analog output voltage of 1.23VDC. When in the 3 T range, a
reading of 1.23T produces the same output, 1.23VDC.
The analog output can reach a maximum output of about 4.25VDC in order to accommodate the peak value of a
3VRMS AC signal. This means that the analog output can be used to measure flux density levels that exceed the
normal range of the displayed readings. For instance, a level of 31.5mT in the 30mT range would normally result in a
flashing 29.99mT overrange condition. However, the output will still be +3.15VDC.
The meter will apply an optimum amount of low pass filtering to the analog output signal based on the frequency of
the detected magnetic field and mode of operation (AC or DC mode).
Figure 3-20
Analog Output Display
3-17