AUTOMATIC RELATIVE MODE
In the automatic relative mode, the present flux density as seen by the probe is used as the relative offset value.
Prepare the probe and select an appropriate range and mode (AC or DC) as needed (AUTO RANGE is
deactivated when RELATIVE MODE is used). To automatically assign the current reading as the relative offset
value, press the REL key.
The AUTO and RELATIVE icons will flash momentarily indicating the operation. When complete, the unit will
return to the measurement mode and the RELATIVE icon will be on indicating that a relative value is being
subtracted from all subsequent readings.
To deactivate the relative mode, press the SHIFT key followed by the REL key.
Difference between ZERO and RELATIVE: ZERO introduces an electrical offset signal into the circuitry.
RELATIVE is strictly a mathematical process.
Figure 3-17
AUTOMATIC RELATIVE Function
3-14