ANALOG OUTPUT FUNCTION
The 5180 meter is capable of providing an analog voltage signal proportional to the
present flux density level. Calibration is set to 3.0V full scale DC or 3.0VRMS AC,
depending upon the mode of operation. This signal, available at the BNC connector,
can be connected to a voltmeter, oscilloscope, recorder, data logger, or external
analog to digital converter.
Power consumption increases when using the analog output. If use of the analog
output port is unnecessary, maximize battery life by disabling the analog output
feature. To toggle the analog output on or off, simply press the SHIFT key followed
by the UP arrow key. The ANALOG ON icon will turn on when the analog output
feature is active.
Figure 3-19
Output Function
3-16