©2022 PACE Inc., Vass, North Carolina, All Rights Reserved Page 15
Display Message Codes
Display Message
Description
The incorrect password has been entered. The displayed message will
time out after 6 seconds and revert to normal operation. Enter the
correct password.
ISB= Instant SetBack – The handpiece connected to the associated
channel has gone into Instant SetBack.
CHP = Check HandPiece – A handpiece is not connected to the Power
Receptacle or is faulty, or the Tip-Heater Cartridge is not properly
inserted or faulty.
OCE = Over Current Error – The Tip-Heater Cartridge circuit is shorted,
or the handpiece has failed. Replace the Tip-Heater Cartridge.
SBk= SetBack – Indicates the unit is in Temperature SetBack mode
and has lowered the temperature to prolong tip life by preventing
oxidation.
Spare Parts
Description
PACE Part Number
Fuse 1.25 Amp, Time Lag (for 120V)
1159-0275-01-P5
Fuse 0.63 Amp, Time Lag (for 230V)
1159-0275-02-P5
MBT Motor/Pump Assembly
1336-0083-P1
Motor/Pump Rebuild Kit
6993-0260-P1
Optional under shelf mounting bracket
1321-0609-P1