OPERATION/MAINTENANCE
Setting tone detection options requires a high level of understanding of fiber optics, and should only be performed
by advanced fiber technicians
From the WAVE OPTIONS menu, press
to view the TONE OPTIONS that are set in the Fiber
OWL 4.
TONE OPTIONS are used to associate five tone detection frequencies with five calibrated
wavelengths, as well as set the Fiber OWL 4 into AUTO WAVELENGTH detection mode.
The screen at right shows the default tone options.
TONE - moves the tone selection highlight to the next tone.
WAVE - toggles the wavelength of the currently selected tone.
AUTO - toggles the AUTO WAVELENGTH detection feature ON or OFF.
NOTE: for proper operation of automatic wavelength detection with OWL WaveSource light sources, the toning options
MUST appear as shown in the Set Toning Options screenshot above.
3
GHI
F1
F2
F3
SETTING TONE DETECTION OPTIONS
From the METER SETUP MENU, press
to set the LCD CONTRAST.
This option allows the user to set the contrast of the liquid crystal display (LCD).
DOWN - lightens the screen in case it is too dark
UP - darkens the screen in case it is too light
Press
to return to the METER CONFIG MENU.
4
JKL
F1
F3
DONE
SETTING LCD CONTRAST
Set Toning Options
NOTE: shown above is the
default tone configuration
TONE
AUTO
600Hz
1000Hz
1500Hz
2000Hz
300Hz
_____________________
WAVELENGTH: OFF
WAVE
AUTO
1310nm
1300nm
1490nm
1550nm
850nm
Set LCD Contrast
XXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXX
XXXXXXXXXXXXXXXXXXXXX
DOWN
XXXXXXXXXXXXXXXXXXXXX
CONTRAST
XXXXXXXXXXXXXXXXXXXXX
_____________________
UP