<1>CUSTOM
<2>DEFAULT
<3>TONE DETECTION
OPTIONS
WAVELENGTH
WAVELENGTH
WAVELENGTH
WAVELENGTH
OPTIONS MENU
STARTUP
STATE:
SDWN
TEMPERATURE
AUTO
_____________________
SHUTDOWN:
BACKLIGHT
ON
F
UNITS:
ON
BKLT
TEMP
Set Power
Saving Features
OPERATION/MAINTENANCE
From the METER SETUP MENU, press
to set User Preferences, which
includes a power saving feature and the displayed temperature of the Fiber OWL
4.
toggles the auto-shutdown feature between ON and OFF. This feature is
ON by default.
toggles between Fahrenheit (F) and Celcius (C) degrees. This feature is
Fahrenheit (F) by default.
2
DEF
F1
F3
SETTING USER PREFERENCES
From the METER SETUP MENU, press
to set the various wavelength-
related options in the Fiber OWL 4, including setting custom wavelengths and
3
GHI
Setting custom wavelengths requires a high level of understanding of fiber optics,
and should only be performed by advanced fiber technicians
From the WAVE OPTIONS menu, press
to set a custom wavelength. The custom
wavelength temporarily replaces 980nm, and requires a singlemode light source tuned to -10dBm
for calibration.
Enter the 3- or 4-digit custom wavelength between 700nm and 1700nm in the entry field provided; e.g. 1650nm.
Press
to continue.
Connect a -10 dBm singlemode light source of the appropriate wavelength to the Fiber OWL 4
using a singlemode patch cord.
Press
to confirm calibration. The meter will then return to the METER SETUP MENU.
To reset the Fiber OWL 4 back to its default wavelengths, press
from the WAVE OPTIONS
1
ABC
DONE
F1
2
DEF
CHANGING WAVELENGTH OPTIONS
ENTERING CUSTOM WAVELENGTH
Set Custom
Wavelength
(
[1650]
WAVELENGTH(nm):
ENTER
_____________________
<---
--->
SHIFT
700 to 1700nm )
980nm
CONNECT
YES
SOURCE
1650nm
_____________________
WILL REPLACE
1650nm -10dBm
TO CALIBRATE
NO
Verify Custom
Wavelength