USER'S MANUAL
General instructions for use
I-Max Touch (110-120V)
(Rev. 0)
76
5.
Instruct the patient to close his eyes.
6.
Press the key "Centring devices ON" (14) .
Two laser beams will light up the sagittal medial plane line and the
horizontal line for the Frankfurt plane reference (the plane that
identifies a line that ideally links the ear hole - the auditory meatus -
with the lower part of the orbital fossa in Figure 10). Position the
patient's head in such a way as to ensure that the first two luminous
beams fall in correspondence with the respective anatomical
references.
The luminous beam of the Frankfurt plane can be adjusted according
to the patient's height; this can be adjusted by means of the laser
knob on the side of the mirror.
NOTE:
The laser centring device will stay lit up until the "Centring Devices ON"
(14) key or, at the end of the alignment, the key "Patient
Entrance" (6) is pressed to start the exposure.
7.
Close the temple clasps and bring the forehead support close; this
will help the patient to stay in a correct position. Check that, during
this phase, the patient has not changed position.
8.
Press the key "Patient Entrance" (6) to confirm.
The luminous centring devices switch off and the rotating arm goes
to its examination start position. Once alignment has been
completed, the following message will be displayed:
x x k V x x m A 9 . 4 0 s
S T A R T E X A M
x = value defined by the settings
9. Ask the patient to: close his mouth, remain perfectly still and
not look at the rotating arm during the movement.
*