USER'S MANUAL
General instructions for use
(Rev. 0)
I-Max Touch (110-120V)
53
6.5.4
Making an exposure
NOTE:
When the key "Test" (5) is pressed the Test function is
activated.
In this condition, it will be possible to make the unit perform all the
movements made during the examination
without emitting X-rays
.
Once the cycle is completed, deactivate the "Test" function by pressing
the key again.
WARNING:
During the emission of X-rays, the protection procedures for the operator
and personnel in the area must be in compliance with the local
regulations.
In all cases, it is recommended that during the emission of X-rays, only
the patient and operator be present in the room. If the operator is not
protected by suitable screens, he must stand at least 1.5 m (4.9 ft) away
from the emission of the rays (see the Figure 1 and Figure 2).
1.
Verify once again that the exposure data are correct. If not, correct
them as described in paragraph 6.5.2.2; ensure that the machine's
indicator light "Ready for X-rays" will come on, so press the X-ray
button for the entire duration of the exposure, checking the
simultaneous working of the X-ray indicator light "X-rays emission"
(if you are within sight of the machine) and the acoustic ray signal.
The following message will be displayed first:
S
T
A
R
T
E
X
A
M
P R E - H E A T I N G . . .
and then (after 2 seconds), the following message will be displayed:
x x k V x x m A x x . x s
>
X
- R
A
Y
<
x = value defined by the settings
NOTE:
If the machine is in the "Test" mode, the display will show:
T E S T
X R A Y N O T A C T I V E
*
*