USER'S MANUAL
General instructions for use
(Rev. 0)
I-Max Touch (110-120V)
101
6.10.3.5 E362 - X-ray button released during examination
NOTE:
The X-ray button has the so-called "dead man’s switch" function, i.e. it
must be kept pressed for the whole time of the examination,
also during
the phases of the examination with emission interruption
(for
instance, in open/close mouth TMJ).
This message signals that the button was released during the
examination phase; the motors are unlocked, therefore the patient can
get out of the system. Repeat the system centring phase and repeat the
examination.
*