Storing of automatical exposure parameters
The pre-set technical exposure factors can be varied according to the
needs of the user, or to obtain somewhat more contrasted images.
To modify the automatic exposure parameters, please follow the indicated
procedure:
1.
Select the examination, the type of patient and the size to be
modified.
2.
Modify the KV, mA and/or time (for Cephalometric programs)
parameters to suit your needs; the "Anatomic/Manual" mode
indicator changes to manual .
New parameters can only be saved in "Manual" mode.
3.
Press the "Anatomic/Manual" mode indicator until it turns green and
displays "S" , then press the "Examination mode Selection"
key (10) to store the modified parameters for the
examination and type and size of patient you have selected.
4.
After pressing the key, the display will show the following message:
S
T
O
R
E
?
> 0 < = Y
, T = N
Press the "Patient Entrance" (6) key to confirm or the key
"Test" (5) key to cancel the setting.