6.20.2 Replacing the control panel
The standard configuration has the control panel mounted on the right side. Depending on wheth-
er the wheelchair user is a right-handed or left-handed, it can be mounted to either arm rest.
NOTICE
Cable damage.
Positioning the cables incorrectly can lead to pinching and damage to the cables. The
cables must not be attached too tightly or too loosely. Avoid bending or squeezing the cables.
INfORMATION
Should the connection plug come loose while driving, e. g. due to vibrations, the connection to the
control panel will be lost and the power wheelchair triggers an emergency stop. When connecting the
plugs, ensure that they are firmly connected to each other.
Tools:
❒
Allen wrench, size 3
Steps:
1. Disconnect the plug connection on the controller.
2. Loosen all cable ties attaching the cable to arm rest.
3. Loosen three set screws on the bottom of the arm rest.
4. Pull the control panel with control panel holder forward out of the track.
5. Attach the control panel with control panel holder on the opposite side of the power wheel-
chair.
6.20.3 Replacing the control panel holder
Preparation:
❒
Remove the control panel with control panel holder (see Section 6.20.2).
Tools:
❒
Phillips screwdriver
Steps:
1. Separate the control panel and control panel holder; in order to do so, loosen four Phillips
head screws on the bottom of the control panel holder.
2. Replace the control panel holder.
Properly reassemble all components upon completion of the work.