![Otto Bock C1000 DS Скачать руководство пользователя страница 107](http://html.mh-extra.com/html/otto-bock/c1000-ds/c1000-ds_service-instructions-manual_1659937107.webp)
107
8.3.2 Electrical specifications
Power supply:
Operating voltage
16 – 33 V
Power supply via bus
12 A
Power supply direct from the battery (continuous)
18 A
Power supply direct from the battery (3 seconds)
24 A
Standby current (typical)
4 mA
Table 5 Power supply
A 2-pin bridge between poles 1 and 5 of the battery connector is required for the power supply
of the seat module via the bus. If more current is required, the seat module can be wired directly
to the battery (poles 1–4). In this case a fuse must be installed as shown in the wiring diagram.
Actuator outputs (A1–A5):
All five actuators can operate forwards or backwards independently in a speed range from
0 to 100%.
Nominal current
1
)
For 1 second
25 A
For 10 seconds
15 A
Continuous (maximum)
10 A
2
)
1) Data apply when only one of the actuators is in operation.
2) This nominal current depends on the seat module installation.
❒
Up to 5 actuators may be operated simultaneously. The total current is not permitted to ex-
ceed the maximum supply current of 18 A.
❒
Each actuator is short-circuit protected (M+ to M-). There is no protection against incorrect
wiring between battery B+ or B- and the actuator connector, so that the module could be
destroyed in this case.
❒
The absolute current limit of the actuators can be set by a parameter. This parameter ap-
plies to all actuators.
❒
The actuator outputs have high impedance when they are not in operation.
INfORMATION
The enAble50 instructions for use 647G490=D/GB contain information on the power supply and ac-
tuator connectors.