NOTICE!
Risk of product damage and limited functionality
►
To prevent mechanical damage, use caution when working with
the product.
►
If you suspect the product is damaged, check it for proper func
tion and readiness for use.
►
Do not use the product if its functionality is restricted. Take suit
able measures (e.g. cleaning, repair, replacement, inspection by
the manufacturer or a specialist workshop).
Signs of changes in or loss of functionality during use
Instability or lack of joint function indicate a loss of function.
4 Scope of delivery
15K10 modular foam kit
Fig.
Item
Quant
ity
Designation
Reference
number
–
–
1(
)
Instructions for use
647G456
1
–
1(
)
Forearm moulded foam
part
–
1
–
1(
)
Upper arm moulded
foam part
–
1
–
1(
)
Perlon frizz stockinette
–
1
–
1(
)
Mounting flange
13R8=64
6
5 Preparing the product for use
5.1 Preparation
>
The modular prosthesis has been prepared.
1)
INFORMATION: In both areas, add an allowance of 2 to 3 cm
for compression.
Forearm area:
determine the length of the prosthesis from the
terminal device to the modular elbow joint.
Upper arm area:
determine the length of the prosthesis from the
modular elbow joint to the prosthesis socket/prosthetic shoulder
joint.
2) Determine the circumferences on the contralateral side. For a
bilateral fitting, pay attention to the overall aesthetic impression.
3) Cut the foam moulded parts to length.
5.2 Assembling
>
Materials needed:
636W17 plastic adhesive, 13R9* connection
disc with groove (for connection with a wrist), 13R8=64 connec
tion flange (for connection to the lamination ring), 623T8=9 Perlon
frizz stockinette
1) Bond the required connection components to the moulded foam
parts with plastic adhesive.
2) Push the moulded foam parts onto the modular prosthesis.
3)
Optional:
Bond the moulded foam parts to each other.
4) Grind the contour of the cosmetic foam cover.
5) Pull a double layer of Perlon frizz stockinette over the cosmetic
foam cover and secure it to the connection disc.
6 Cleaning
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
7 Maintenance
►
A visual inspection and functional test of the prosthetic compo
nents should be performed after the first 30 days of use.