13 Technical data
Water level
Measuring range
Variant 1 0 … 10 m ·33 ft water column /0 ...1 bar
Variant 2 0 … 20 m ·66 ft water column /0 ... 2 bar
Variant 3 0 … 40 m ·131 ft water column /0 ... 4 bar
Variant 4 0 … 100 m ·328 ft water column /0 ... 10 bar
Resolution 0.001 m · 0.1 cm · 1 mm
0.001 inch · 0.001 ft
0.00001 bar · 0.01 mbar
0.0001 psi
Accuracy (lin hysteresis)
Standard specification ≤ ± 0.05 % full scale
USGS specification (valid for variant code 1) 0 … 527 mbar: ≤ ±0,3 mbar
527 … 1000 mbar: ≤ ±0,5 mbar
Long-term stability (lin hysteresis) ≤ ±0.1 %/a full scale
Zero drift ≤ ±0.05 % full scale
Units m · cm · mm
inch · ft
bar · mbar · kPa
psi
Overload protection for the measuring cell
(without permanent mechanical damage)
Variant 1 (0 … 10 m · 33 ft WC / 0 ... 1 bar) 6.7 bar
Variant 2 (0 … 20 m · 66 ft WC / 0 ... 2 bar) 12 bar
Variant 3 (0 … 40 m · 131 WC / 0 ... 4 bar) 16.7 bar
Variant 4 (0 … 100 m · 328 ft WC / 0 ... 10 bar) 26.7 bar
Pressure sensor ceramic (Al
2
O
3
99,9 %), capacitive
Temperature-compensated operating range –20 (ice free) … +70 °C
· –4 (ice free) … +158 °F
Water temperature
Measuring range –40 … +70 °C
· –40 … +158 °F
Resolution 0.01 °C · 0.01 °K
Accuracy
–20 … +50 °C · – 4 … +122 °F typ. ±0.05 °C; max. ±0.1 °C · typ. ±0.09 °F; max. ±0.18 °F
–40 … +70 °C · –40 … +158 °F typ. ±0.05 °C; max. ±0.15 °C · typ. ±0.09 °F; max. ±0.27 °
Units °C, °K, °F
Relative humidity in the probe housing
Measuring range 0 … 100 % rH
Resolution 1 % rH
Accuracy typ. ±2 % rH (20 … 80 % rH)
max. ±3 % rH (0 … 100 % rH)
Units
Supply voltage +5.5 … +28.8 V
DC
; typ. 12/24 V
DC
Current consumption
idle < 250 μA; typ. 15 μA
active < 4 mA; typ. 2.9 μA
Interfaces SDI-12 version 1.4
EIA-485 (RS-485); SDI-12 protocol
Modbus RTU
Measured physical parameters – water level/hydrostatic pressure
– water temperature
– relative humidity in the probe housing
– sensor orientation