Command
Response
Description
aRC0!
a<value1><value2><value3>
…
…
<value4>
<CRC><CR><LF>
The OTT PLS 500 continuously measures level/pressure and
temperature, optionally calculates the discharge and
requests a CRC value (Cyclic Redundancy Check). This com-
mand permits to retrieve measurement results, even without
the combination of commands
aM!
/
aD0!
; for details, see
command
aD0!
after
aM!
. Requires measurement type
"continuous measurement"; extended command
aXXC<value>!
aM1!
atttn<CR><LF>
and after 0/1 … 60 seconds
a<CR><LF>
Start measurement – including statistical values
a
– sensor address
ttt
– time in seconds until the sensor has determined
the measurement result
response OTT PLS 500: 0/1 … 60 seconds
1)
n
– number of measured values
response OTT PLS 500:
8
aD0!
a<value1><value2><value3>
…
…
<CR><LF>
Send data (after
aM1!
)
a
– sensor address
<value1>
– last single measured value "level/pressure"
within the averaging time
2)
<value2>
– water temperature
2)
<value3>
– mean of measured values "level/pressure"
within the averaging time
2) 3)
aD1!
a<value4><value5><value6>
…
…
<CR><LF>
Send data (after
aM1!
)
a
– sensor address
<value4>
– minimum of measured values "level/pressure"
within the averaging time
2) 3)
<value5>
– maximum of measured values "level/pressure"
within the averaging time
2).3)
<value6>
– median of measured values "level/pressure"
within the averaging time
2) 3)
aD2!
a<value7><value8><CR><LF>
Send data (after
aM1!
)
a
– sensor address
<value7>
– standard deviation of measured values
"level/pressure" within the averaging time
2).3)
<value8>
– device status; see
aD2!
after
aV!
aMC1!
atttn<CR><LF>
and after 0/1 … 60 seconds
a<CR><LF>
Start measurement and request CRC (Cyclic Redundancy
Check); for details, see command
aM1!
.
The response to the following
aD0!
…
aD2!
command is
extended by a CRC value:
a<valueX><valueX><valueX><CRC><CR><LF>
aC1!
atttnn<CR><LF>
Start concurrent measurement (simultaneous measurement
with multiple sensors on one single bus line); for details, see -
command
aM1!
. The number of measured values in the
response to this command has two digits:
nn
=
08
.
aCC1!
atttnn<CR><LF>
Start concurrent measurement (simultaneous measurement
with multiple sensors on one single bus line) and request
CRC (Cyclic Redundancy Check); for details, see command
aM1!
. The number of measured values in the response to
this command has two digits:
nn
=
08
. The response to the
following
aD0!
…
aD2!
command is extended by a CRC
value:
a<valueX><valueX><valueX>
…
…
<CRC><CR><LF>
1)
depending on the set averaging time; extended command
aXXM<value>!
In the measurement type "continuous measurement" the time is always 0 seconds (with the exception of the first interval after the start)!
2)
for measured value formats, see
aD0!
after
aM!
(depending on the set unit)
3)
determined from 2 … 238 single measurements of a measurement interval (see command
aXXC!
)
26