Command
Response
Description
aR1!
a<value1><value2><value3>
…
…
<value4><value5><value6>
…
…
<value7><value8><CR><LF>
The OTT PLS 500 continuously measures level/pressure and
temperature and determines statistical values. This command
permits to retrieve measurement results even without the com-
bination of commands
aM1!
/
aD0!
…
aD2!
; for details, see
command
aD0!
…
aD2!
after
aM1!
.
Requires measurement type "continuous measurement";
extended command
aXXC<value>!
aRC1!
a<value1><value2><value3>
…
…
<value4><value5><value6>
…
…
value7><value8><CRC><CR><LF>
The OTT PLS 500 continuously measures level/pressure, tem-
perature, determines statistical values and requests a CRC
value (Cyclic Redundancy Check). This command permits to
retrieve measurement results even without the combination of
commands
aM1!
/
aD0!
…
aD2!
; for details, see command
aD0!
…
aD2!
after
aM1!
. Requires measurement type
"continuous measurement"; extended command
aXXC<value>!
aM2!
atttn<CR><LF>
and after 0/1 … 60 seconds
a<CR><LF>
Start measurement – including meta data
a
– sensor address
ttt
– time in seconds until the sensor has determined
the measurement result
response OTT PLS 500: 0/1 … 60 seconds
1)
n
– number of measured values
response OTT PLS 500:
9
aD0!
a<value1><value2><value3>
…
…
<CR><LF>
Send data (after
aM2!
)
a
– sensor address
<value1>
– relative humidity in the probe housing
measured value format:
pbb.ee
[
% rH
]
<value2>
– dew point in the probe housing
measured value format:
pbb.ee
[°C]
<value3>
– current sensor orientation
measured value format:
pbbb
[°]
aD1!
a<value4><value5><value6>
…
…
<CR><LF>
Send data (after
aM2!
)
a
– sensor address
<value4>
– stored sensor orientation during installation
measured value format:
pbbb
[°]
<value5>
– temperature value of pressure sensor
measured value format:
pbb.ee
[°C]
<value6>
– temperature of internal humidity sensor
measured value format:
pbb.ee
[°C]
aD2!
a<value7><value8><value9>
…
…
<CR><LF>
Send data (after
aM2!
)
a
– sensor address
<value7>
– pressure value of pressure sensor
measured value format:
pbbbb.ee
[mbar]
<value8>
– standard deviation of pressure value of
pressure sensor
2)
measured value format:
pbbbb.ee
[mbar]
<value9>
– device status; see
aD2!
after
aV!
aMC2!
atttn<CR><LF>
and after 0/1 … 60 seconds
a<CR><LF>
Start measurement and request CRC (Cyclic Redundancy
Check); for details, see command
aM2!
.
The response to the following
aD0!
…
aD2!
command is extended by a CRC value:
a<valueX>
<valueX><valueX><CRC><CR><LF>
1)
depending on the set averaging time; extended command
aXXM<value>!
In the measurement type "continuous measurement" the time is always 0 seconds (with the exception of the first interval after the start)!
2)
determined from 2 … 238 single measurements of a measurement interval (see command
aXXC!
)
27