Advanced commands (manufacturer-specific)
▶
aXSU<value>!
Set the unit for level/depth measurements
aXSU!
Read the unit for level/depth measurements
Factory setting: +
0
➝
m (presetting metric); +
3
➝
ft (presetting imperial)
▶
aXST<value>!
Set the unit for temperature values
aXST!
Read the unit for temperature values
Factory setting: +
0
➝
°C (presetting metric); +
1
➝
°F (presetting imperial)
▶
aXSD<value>!
Set the unit for discharge values
aXSD!
Read the unit for discharge values
Factory setting: +
0
➝
m
3
/s (presetting metric); +
2
➝
ft
3
/s (presetting imperial)
▶
aXXG<value>!
Set local gravitational acceleration
aXXG!
Read local gravitational acceleration
Factory setting:
+9.806650
m/s
2
▶
aXXS<value>!
Set salinity
aXXS!
Read salinity
Factory setting:
+0.000000
mg/l
▶
aXXR<value>!
Set average water density
aXXR!
Read average water density
Factory setting:
+0.999975
kg/dm
3
▶
aXXO!
Store sensor orientation during installation
▶
aXAA<value>!
Set measuring mode "level" or "depth"
aXAA!
Read measuring mode
Factory setting:
+0
➝
Measuring mode "level measurement"
▶
aXXM<value>!
Set averaging time
aXXM!
Read averaging time
Factory setting:
+1.5
s
▶
aXXC<value>!
Set measuring mode
aXXC!
Read measuring mode
Factory setting:
+0
➝
Measuring mode "single measurement"
▶
aXAB<value>!
Set offset value for level/depth measurements
aXAB!
Read offset value
Factory setting:
+0.000
m
▶
aXAC<value>!
Set reference value for level/depth measurements
aXAC!
Read reference value
Factory setting:
+0.000
m
▶
aXSR<value>!
Reset factory settings for units (metric or imperial)
aXSR!
Read factory settings for units
Factoring setting: depending on the ordered variant code
▶
aXSF!
Reset pressure probe to factory settings without communication settings
aXSF+1!
Reset pressure probe to factory settings including communication settings
▶
aXDC<value>!
Set calculation method for discharge measurement
aXDC!
Read calculation method for discharge measurement
Factory setting:
+0
➝
discharge measurement "deactivated"
▶
aXDA<value1><value2>!
Create table entry rating table (calculation method rating table)
▶
aXDA<value1><value2><value3>!
Enter coefficients for discharge measurement (exponential formula)
▶
aXDR<value>!
Read table entry rating table (calculation method rating table)
aXDR!
Read number of entries in rating table (calculation method rating table)
aXDR!
Read coefficients for discharge measurement (calculation method exponential formula)
▶
aXDD<value>!
Delete table entry rating table
aXDD+9999!
Delete rating table completely
22