Command
Response
Description
aD0!
a<value1><value2><value3>
…
…
<value4>
<CR><LF>
Send data (after
aM!
)
a
– sensor address
<value1>
– mean of measured values "level/pressure"
within the averaging time
measured value formats
1)
:
pbbb.eee
[m]
pbbbbb.e
[cm]
pbbbbb
[mm]
pbbb.eee
[ft]
pbbbb.eee
[inch]
pbb.eeeee
[bar]
pbbbb.ee
[mbar]
pbbbb.eee
[kPa]
pbbb.eeee
[psi]
<value2>
– water temperature
measured value formats
2)
:
pbb.ee
[°C]
pbbb.ee
[°K]
pbbb.ee
[°F]
<value3>
– device status
see
aD2!
after
aV!
<value4>
– discharge
4)
measured value formats
3)
:
pbbb.eee
[m
3
/s]
pbbbbbb
[l/s]
pbbbbb.eee
[ft
3
/s]
Note
Measured value =
-9999
➝
calculation error occurred or
rating table missing; =
-9998
➝
entries in the rating table
are not sufficient for the calculation
aMC!
atttn<CR><LF>
and after 0/1 … 60 seconds
a<CR><LF>
Start measurement and request CRC (Cyclic Redundancy
Check); for details, see command
aM!
.
The response to the following
aD0!
command is extended
by a CRC value:
a<value1><value2><value3>
…
…
<value4>
<CRC><CR><LF>
aC!
atttnn<CR><LF>
Start concurrent measurement (simultaneous measurement
with multiple sensors on one single bus line); for details, see -
command
aM!
. The number of measured values in the
response to this command has two digits:
nn
=
03
or
04
4)
.
aCC!
atttnn<CR><LF>
Start concurrent measurement (simultaneous measurement
with multiple sensors on one single bus line) and request CRC
(Cyclic Redundancy Check); for details, see command
aM!
.
The number of measured values in the response to this com-
mand has two digits:
nn
=
03
or
04
4)
.
The response to the following
aD0!
command is extended by
a CRC value:
a<value1><value2><value3>
…
…
<value4>
<CRC><CR><LF>
aR0!
a<value1><value2><value3>
…
…
<value4>
<CR><LF>
The OTT PLS 500 continuously measures level/pressure,
temperature and optionally calculates the discharge. This
command permits to retrieve measurement results even without
the combination of commands
aM!
/
aD0!
; for details, see
command
aD0!
after
aM!
. Requires measurement type
"continuous measurement"; extended command
aXXC<value>!
1 … 3)
depending on the set unit; extended command
aXSU<value>!
1)
;
aXST<value>!
2)
;
aXSD<value>!
3)
4)
optional with activated discharge measurement; exteded command
aXDC<value>!
25