Measured value overview standard commands
1)
metric units imperial units
▶
Send data (
D0
) after
aM!
command
<value1>
level/pressure
pbbb.eee
[m]
pbbb.eee
[ft]
<value2>
water temperature
pbb.ee
[°C]
pbbb.ee
[°F]
<value3>
device status see below
<value4>
discharge
2)
pbbb.eee
[m
3
/s]
pbbbbb.eee
[ft
3
/s]
▶
Send data (
D0
,
D1
,
D2
) after
aM1!
command
<value1>
last single measured value "level/pressure"
pbbb.eee
[m]
pbbb.eee
[ft]
within the averaging time
<value2>
water temperature
pbb.ee
[°C]
pbbb.ee
[°F]
<value3>
mean of measured values "level/pressure"
pbbb.eee
[m]
pbbb.eee
[ft]
within the averaging time
<value4>
minimum of measured values "level/pressure"
pbbb.eee
[m]
pbbb.eee
[ft]
within the averaging time
<value5>
maximum of measured values"level/pressure"
pbbb.eee
[m]
pbbb.eee
[ft]
within the averaging time
<value6>
median of measured values "level/pressure"
pbbb.eee
[m]
pbbb.eee
[ft]
within the averaging time
<value7>
standard deviation of measured values
pbbb.eee
[m]
pbbb.eee
[ft]
"level/pressure" within the averaging time
<value8>
device status; see below
▶
Send data (
D0
,
D1
,
D2
) after
aM2!
command
<value1>
– relative humidity in the probe housing
pbb.ee
[
% rH
]
pbb.ee
[
% rH
]
<value2>
– dew point in the probe housing
pbb.ee
[°C]
pbb.ee
[°C]
<value3>
– current sensor orientation
pbbb
[°]
pbbb
[°]
<value4>
– stored sensor orientation during installation
pbbb
[°]
pbbb
[°]
<value5>
– temperature value of pressure sensor
pbb.ee
[°C]
pbb.ee
[°C]
<value6>
– temperature of internal humidity sensor
pbb.ee
[°C]
pbb.ee
[°C]
<value7>
– pressure value of pressure sensor
pbbbb.ee
[mbar]
pbbbb.ee
[mbar]
<value8>
– standard deviation of pressure value of pressure sensor
pbbbb.ee
[mbar]
pbbbb.ee
[mbar]
<value9>
– device status; see below
▶
Device status
3)
+0
➝
no error occured
+1
➝
system reset – status flag is set when connecting the operating voltage/upon unexpected reset;
deletion after status read out
+2
➝
raw value "pressure" outside the calibrated range
(1 % tolerance of full scale is considered)
+4
➝
raw value "temperature" outside the calibrated range
+8
➝
sensor orientation change (deviation since installation currently ≥ 5°)
4)
+16
➝
overload pressure sensor (≥ +20 % of measuring range + offset)
+32
➝
pressure probe was reset to factory settings due to internal system error
(including potential flow settings)
+64
➝
internal relative humidity was/is above limit (≥ 25 % rH)
5)
1)
with factory setting
2)
optional with activated discharge measurement; extended command
aXDC<value>!
3)
if several errors/events occur at the same time, the OTT PLS 500 adds up the status values. Example:
+20
➝
overload pressure sensor (
+16
) + raw
value "temperature" outside the calibrated range (
+4
); values ≥
+128
: exclusively for internal service purposes
4)
change sensor orientation intended (e.g. new installation):
➝
first reset pressure probe to factory setting with command
aXSF!
5)
see also "Please note: …" in Chapter 9
20