8 RS-485 Interface with Modbus protocol (RTU)
8.1 Preconditions
▶
OTT PLS 500 variant code: M
➝
Protocol RS-485 interface Modbus
▶
Measurement type continuous measurement (interval mode or floating mode)
▶
Interface
EIA-485 (RS-485)
▶
Transmission parameter 8 Data bit, 1 Stop bit, Even parity
▶
Transmission speed 9 600 (factory setting), 19200
▶
Bus address 1 … 247
8.2 Value ranges
Note:
The OTT PLS 500 has only one "Holding Register" Block.
Modbus Register
1
Byte
0
1
Bit
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
16-bit Integer values
int range: –32767 … 32767
uint range: 0 … 65534
bitfield16 range: 0 … 0x7FFF
int range: –214483647 … 214483647
uint range: 0 … 4294967294
float32 range: see IEEE 754
32-bit Integer values
Floating point values
String values
Modbus Register
1
Byte
0
1
Bit
31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16
IEEE 754
sign
Exponent
Fraction
Modbus Register
1
2
Byte
0 1 3 4
Bit
31 … 24 23 … 16
15… 08
07 … 00
Modbus Register
2
Byte
2
3
Bit
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
IEEE 754
Fraction least
Modbus Register
1 2 3 4 5 6 7 8
Byte
0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15
Bit E
X
A
M
P
L
E
spc
S
T
R
I
N
G
!
NULL
39