Electrical data
Supply voltage through USB port
Input power
USB 2.0, Max. 500 mA.
Light source
2 LEDs, fibre-optic light guide
USB plug
USB, type A (LED power supply/Camera/Pre-heating)
Switch activated by magnet contact used to switch between pre-heating and light.
Mechanical data
Cable length
2950mm (9.65 feet)
Length without cable
170mm ( 6.7 inches)
Weight including cable
225 g (8 ounces)
Largest diameter
45 mm (1.8 inches)
Distal diameter
max. 3.4 mm (0.134 inches)
Total weight
1300 g
Service life
Expected Service life
5 years
Storage environment
Temperature
-20°C to +60°C (-4°F to +140°F)
Operating environment
Temperature
+10°C to +30°C (+50°F to +86°F)
Air humidity
30% to 75%, non-condensing
Air pressure
600 hPa to 1060 hPa
Warm-up time
Warm-up time
<20 seconds
Note
•
In order for the pre-heater to heat the camera tip to body temperature, AURICAL OTOcam 300 should
remain in the cradle (connected to a PC which is powered on) for minimum 5 minutes before the camera is used.
The time should be extended if AURICAL OTOcam 300 has been stored in a cold environment.
Essential performance
AURICAL OTOcam 300 has no essential performance.
26
Otometrics - AURICAL OTOcam 300
App. 1 Technical Specifications